I5S Fluke Electronics, I5S Datasheet

no-image

I5S

Manufacturer Part Number
I5S
Description
CURRENT CLAMPS AC 5A
Manufacturer
Fluke Electronics
Type
AC Clampr
Datasheet

Specifications of I5S

Style
Pincer
Ac Current Range
10mA ~ 6A
Bandwidth
5kHz
Cable Length
98.425" (2500.00mm)
Max Conductor Size
0.591" (15.00mm)
Termination
BNC
Ratings
CAT III 600V
Lead Free Status / RoHS Status
Lead free by exemption / RoHS compliant by exemption
Other names
2508467
Introduction
The i5s (hereafter called the “Current Clamp”) is compatible with any instrument that can accept
a standard BNC connector and is capable of ac millivolt measurements; the Current Clamp can
also be used with digital multimeters using a BNC-to-dual banana plug adapter. The Current
Clamp produces an output of 400 mV ac per 1 A ac input level.
Contacting Fluke
To contact Fluke, call one of the following telephone numbers:
Or, visit Fluke's Web site at www.fluke.com.
To register your product, visit
Safety Information
To ensure safe operation and service of the Current Clamp, follow these instructions:
Symbols
PN 2458104 May 2005
© 2005 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies.
-
T
W
X
)
P
Read the operating instructions before use and follow all safety instructions.
Use the Current Clamp only as specified in the operating instructions, otherwise the
clamp’s safety features may not protect you.
Adhere to local and national safety codes. Individual protective equipment must be
used to prevent shock and arc blast injury where hazardous live conductors are
exposed.
Do not hold the Current Clamp anywhere beyond the tactile barrier, see Figure 1.
Before each use, inspect the Current Clamp. Look for cracks or missing portions of the
clamp housing or output cable insulation. Also look for loose or weakened
components. Pay particular attention to the insulation surrounding the jaws.
Never use the Current Clamp on a circuit with voltages higher than 600 V CAT III.
CAT III equipment is designed to protect against transients in equipment in fixed
equipment installations, such as distribution panels, feeders and short branch circuits,
and lighting systems in large buildings.
The Clamp is designed for use on insulated conductors only. Use extreme caution when
working around bare conductors or bus bars. Contact with the conductor could result in
electric shock.
Use caution when working with voltages above
60 V dc or 30 V ac rms. Such voltages pose a shock hazard.
Do not apply around or remove from HAZARDOUS LIVE conductors.
Product is protected by double insulation.
Risk of Danger. Important information. See Instruction Sheet.
Risk of Electric Shock.
Canadian Standards Association has certified that the product meets applicable U.S.
and Canadian Standards.
Conforms to relevant European Union directives.
USA: 1-888-44-FLUKE (1-888-443-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
USA Service: 1-888-99-FLUKE (1-888-993-5853)
XW Read First: Safety Information
register.fluke.com
i5s AC Current Clamp
Instruction Sheet
®
Safety Specifications
Category Rating: CAT III 600 V per IEC/EN61010-1, Pollution Degree 2
EMC: EN 61326-1, FCC for emission and immunity
Electrical Specifications
Reference Conditions: 23
no DC component; no adjacent conductor.
Measurement Range: 10 mA to 6 A
Output: 400 mV/A
Accuracy (48 Hz to 65 Hz):
10 mA to 1 A 1% + 5 mA
1 A to 5 A
Phase Shift (48 Hz to 65 Hz):
10 mA to 100 mA
100 mA to 5 A
Crest Factor: ≤ 3, add 0.7 % to accuracy
Typical Bandwidth: 40 Hz to 5 kHz
Working Voltage: 600 V ac rms, in compliance with EN61010
Common Mode Voltage: 600 V ac rms from earth ground, in compliance with EN61010
Input Load Impedance (of host instrument): >1 M
Maximum Non-destructive Current: 70 A
Duty Cycle: 0.01 A to 6 A continuous
Influence of Adjacent Conductor: ≤ 15 mA/A (@ 50/60 Hz)
Influence of Conductor Position in Jaw Opening: 0.5 % of reading (@ 50/60 Hz)
General Specifications
Output Cable Length: 2.5 m
Maximum Conductor Size: 15 mm
Storage Temperature: -20 C to 60 C
Operating Temperature: 0 C to 50 C
Relative Humidity: 10 C to 30 C: 85 %
Temperature Coefficient: 0.01 % X (specified accuracy)/ C
(< 18 C or > 28 C)
Altitude: Operating: 2000 m; Non-operating: 12000 m
Dimensions: 116 x 43 x 23 mm
Weight: 200 g
Instrument Compatibility
The Current Clamp is compatible with any Fluke Multimeter or any other current measurement
device that has the following features:
Measurement Considerations
Observe the following guidelines when making measurements:
Operation
To use the Current Clamp, refer to “Measurement Considerations” and Figure 1:
1.
2.
3.
4.
5.
6.
)
P
BNC or Banana inputs
Input accuracy of 1 % or better to take full advantage of the accuracy of the Current Clamp.
Input impedance of > 1 M
To avoid shock or personal injury, keep fingers behind the tactile barrier, see Figure 1.
Current input impedance on the Fluke 430 Series Three-Phase Power Quality Analyzer is
< 1 MΩ, but it has a special calibration setting for the i5s to achieve full accuracy.
Center the conductor inside the Current Clamp jaw.
Make sure the Current Clamp is perpendicular to the conductor.
For optimal reading, make sure the conductor is centered in the jaws of the Current
Clamp.
Avoid taking measurements close to other current-carrying conductors.
If necessary, add a dual banana plug to BNC adapter to the i5s BNC connector.
Connect the clamp cable to the instrument. If using an adapter, use the common and
volts inputs of the meter and switch it on.
Make sure the measurement instrument is set to the proper range.
Make sure the arrows on each side of the Current Clamp face towards the load of the
circuit.
Open the Current Clamp jaw by pushing the release button toward the jaw until the latch
has cleared the barrier, then press down to open the jaw.
Connect the Current Clamp jaws around the insulated conductor to be measured and
release the button. Ensure that the release button has returned to the original position to
resume the circuit testing.
Complies with U.S. and Canadian Standards CAN/CSA C22.2 No. 61010-1-04 and
No. 61010-2-032-04; UL61010-1
IEC 61010-1 2
1%
4 °
Unspecified
30 C to 40 C: 75 %
40 C to 50 C: 45 %
nd
Edition IEC 61010-02-032
5 C, 20 to 75 % RH; conductor centered in jaw opening;
in parallel with up to 47 pF
XW Warning
Note
in parallel with up to 47 pF
Maintenance
To avoid possible electric shock or personal injury:
If the Current Clamp does not work or perform properly, use the following steps to help isolate
the problem:
1. Inspect the jaw mating surface for cleanliness. If any foreign material is present, the jaw will
2. Verify that the function selection and range on the Multimeter are correct and adjusted to the
Cleaning
Periodically wipe the case with a damp cloth and mild detergent.
Open the jaws and wipe the magnetic pole pieces with a lightly oiled cloth. Do not allow rust or
corrosion to form on the magnetic core ends.
This Fluke product will be free from defects in material and workmanship for one year from the
date of purchase. This warranty does not cover fuses, disposable batteries, or damage from
accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or
handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf. To obtain
service during the warranty period, contact your nearest Fluke authorized service center to obtain
return authorization information, then send the product to that Service Center with a description
of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR
A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING
FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or
limitation of an implied warranty or of incidental or consequential damages, this limitation of
liability may not apply to you.
11/99
Before each use, inspect the Current Clamp. Look for cracks or missing portions of the
clamp housing and output cable insulating cover and for loose or weakened components.
Pay particular attention to the insulation surrounding the clamp jaws.
Do not use a damaged Current Clamp. If a clamp is damaged, tape it shut to prevent
unintended operation. A damaged clamp under warranty will be promptly repaired or
replaced (at Fluke’s discretion) and returned at no extra charge.
not close properly and measurement errors will result.
sensitivity of the Current Clamp.
To avoid damaging the Current Clamp, do not use abrasives or solvents to
clean the clamp.
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Figure 1. i5s Setup
Single Insulated
Current Carrying
Conductor
Release
Button
XW Warning
W Caution
BNC Connector
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
Tactile Barrier
Load Direction
Arrow
Banana
Plug
cab01f.eps