1620-4-N Leader Electronic Co., Ltd., 1620-4-N Datasheet

no-image

1620-4-N

Manufacturer Part Number
1620-4-N
Description
Pressure Sensor
Manufacturer
Leader Electronic Co., Ltd.
Datasheet
Model 1620
Disposable Blood Pressure Sensor
AAMI Specification
Low Cost
Multiple Configurations
1-48
FEATURES
STANDARD RANGE
-30 to 300
Low Cost Disposable Design
Solid State Piezoresistive Sensor
Top Side Pressure Entry
Compatible with Automated Assembly
Equipment
Integral Dielectric Gel Barrier
Fully Tested and Compensated
mmHg
DESCRIPTION
The Model 1620 is a fully piezoresistive silicon pressure sensor
for use in invasive blood pressure monitoring. The sensor is
designed to be used with automated assembly equipment and
can be dropped directly into a customer’s disposable blood
pressure housing. The sensor is designed to meet the
requirements as described in the Association for the
Advancement of Medical Instrumentation (AAMI) specification
for Blood Pressure Transducers.
mounted on a ceramic substrate. Thick-film resistors on the
ceramic substrate are laser-trimmed for compensation and
calibration.
an easy method of attachment to the customers assembly and
protection for the sensing element. A dielectric gel is placed
over the sensor to provide electrical and fluid isolation.
in a 10x12 element array on a ceramic substrate (120 units per
substrate). The products are shipped in anti-static shipping
containers. Performance characteristics and packaging can be
easily tailored on a special order basis to meet the requirements
of specific customers.
The pressure sensor consists of a pressure sensing element
A plastic cap is attached to the ceramic substrate to provide
The Model 1620 pressure sensors are batch manufactured
Disposable Blood Pressure
Measurements
Kidney Dialysis Machines
Medical Instrumentation
Blood Pressure

Related parts for 1620-4-N

1620-4-N Summary of contents

Page 1

... Measurements Kidney Dialysis Machines Medical Instrumentation DESCRIPTION The Model 1620 is a fully piezoresistive silicon pressure sensor for use in invasive blood pressure monitoring. The sensor is designed to be used with automated assembly equipment and can be dropped directly into a customer’s disposable blood pressure housing. The sensor is designed to meet the ...

Page 2

... VDC Dielectric Gel - 0.5 grams resistor 5. One discharge per minute. 6. Sterilization performed by customer. 7. Defined as common mode symmetry between signal output and either excitation terminal. Model 1620 MAX UNITS NOTES 300 mmHg PSI ±20 mmHg 5.05 µV/V/mmHg 102.5 mmHg 1 1 mmHg ...

Page 3

... Model 1620 CONNECTIONS AND DIMENSIONS All dimensions are in inches [millimeters]. BODY STYLE N BODY STYLE W SECTION A-A 1-50 Disposable Blood Pressure Sensor *Dimensions indicated do not include flares. *Dimensions indicated do not include flares. Notes 1. Materials Used: Substrate: 96% Alumina Transducer (Die): Silicon Die Attach Adhesive: Room Temperature Vulcanizer ...

Page 4

... Each plate may include units that have failed visual or electrical parameters as well as good units. Bad units are identified with a dot on the backside of the cell location. 3. Plates are shipped in dust free anti-static containers to prevent contamination of the gel surface. SCHEMATIC DIAGRAM Blood Pressure ORDERING INFORMATION 1620 - Body Style (N = Narrow Wide) Pad Configuration ( Model Model 1620 PRELIMINARY April 2000 ...