IRC350 EXTECH [Extech Instruments Corporation.], IRC350 Datasheet

no-image

IRC350

Manufacturer Part Number
IRC350
Description
Infrared Calibrator
Manufacturer
EXTECH [Extech Instruments Corporation.]
Datasheet
User Guide
Infrared Calibrator
Model IRC350

Related parts for IRC350

IRC350 Summary of contents

Page 1

... User Guide Infrared Calibrator Model IRC350 ...

Page 2

... Introduction Congratulations on your purchase of the Extech IRC350 Infrared Calibrator. This easy to use portable calibrator can be used to accurately calibrate Infrared (IR) thermometer devices. This instrument is shipped fully tested and calibrated and, with proper use, will provide years of reliable and accurate calibration of temperature sensors and devices. ...

Page 3

... Although the instrument has been designed for optimum durability and trouble-free operation, there are some precautions that should be taken. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and cleaning recommendations can be found in the Maintenance section of this User Guide. than 212°F (100°C). 3 Model IRC350 Version 4.0 06/08 ...

Page 4

... Temperature differences between the IR surface and the reference probe are calculated as shown in the graph below: IRC-350 Gradient between probe and IR surface 0 -0.1 -0.2 -0 -0.4 -0.5 -0.6 50 100 150 200 250 300 350 C (reference Probe) ±1.2°F/0.5°C ±0.2°F/0.1°C 10%), 3A 4 Model IRC350 Version 4.0 06/08 ...

Page 5

... DOWN –Used to decreases the set-point value and also to select the desired unit of measure (ºF or ºC) • UP – Used to increases the set-point temperature and also to select the desired unit of measure (ºF or ºC) 5. Target assembly 6. Test aperture for placing a temperature sensor. 5 shows the desired (set-point) 5 Model IRC350 Version 4.0 06/08 ...

Page 6

... Slots are provided for airflow. The area around the calibrator must be kept clear to allow adequate ventilation. The airflow is directed out the front and can be extremely hot. FUSE FUSE ON OFF 6 Model IRC350 Version 4.0 06/08 ...

Page 7

... Press the “SET” button to store the change. The green LED display will stop flashing. 5. Press “SET” again if both temperature displays are not present. ” buttons at the same time. 7 Model IRC350 Version 4.0 06/08 ...

Page 8

... For the latest version of this User’s Guide, Software updates, and other up-to-the-minute product information, visit our website: www.extech.com Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451 Copyright © 2008 Extech Instruments Corporation (a FLIR company) All rights reserved including the right of reproduction in whole or in part in any form. 8 Model IRC350 Version 4.0 06/08 ...