GSM7-15 SL Power Electronics Manufacture of Condor/Ault Brands, GSM7-15 Datasheet

no-image

GSM7-15

Manufacturer Part Number
GSM7-15
Description
PS MEDICAL SWITCHER 15 V .47 A
Manufacturer
SL Power Electronics Manufacture of Condor/Ault Brands
Series
GSM7r
Datasheet

Specifications of GSM7-15

Rohs Status
RoHS non-compliant
Voltage - Output
15V
Number Of Outputs
1
Power (watts)
7W
Applications
Medical
Power Supply Type
Switching (Open Frame)
Voltage - Input
85 ~ 264VAC
Mounting Type
Chassis Mount
1st Output
15 VDC @ 500mA
Size / Dimension
2.28" L x 2" W x 0.56" H (57.9mm x 50.8mm x 14.3mm)
Power (watts) - Rated
7W
Operating Temperature
0°C ~ 50°C
Efficiency
72%
Approvals
CE, EN, UL
Load Regulation
0.75%
3rd Output
-
2nd Output
-
4th Output
-
Other names
271-1102
14
GSM7 Medical
7 Watt Single Output Global Performance Switchers
SPECIFICATIONS:
Ac Input
85-264 Vac, 47-63 Hz single phase.
Holdup Time
16 ms minimum from loss of ac.
Output Power
Normal continuous output power is 7 W, 9 W peak for 60 s
maximum duration, 10% duty cycle. Factory set to begin
power limiting at approximately 9 W.
Output Regulation
Regulation from initial setpoint measured by changing
load from no-load to 50% load or 50% load to full load in
either direction. Initial sepoint tolerance is measured at
50% load. No minimum load is required to maintain
proper regulation.
Overload Protection
Fully protected against short circuit and output overload.
Short circuit protection is cycling type power limit.
No Load Turn-on/Standby
Full regulation at no load.
Output Noise
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential
mode. Measured with scope probe directly across output
terminals of the power supply with load terminated with
0.1 uF capacitor.
Voltage Adjustment
Factory set with fixed resistors to maximize reliability.
Efficiency
72% minimum for the 5.0 V model at full rated load,
nominal input voltage. Efficiency increases as output
voltage increases.
Input Protection
Internal ac fuse provided on all units. Designed to open
only if catastrophic failure occurs in the unit; fuse does not
blow on overload or short circuit. Internal 320 Vac MOV
included. Units will not be damaged when tested for input
transients in accordance with the following standards:
ESD—IEC1000-4-2 Level III
Fast Transient—IEC1000-4-4 Level III
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com
Environmental
Designed for 0 to 50°C operation at full rated output
power; derate output current and total output power by
2.5% per °C above 50°C. See Environmental specifica-
tion on next page.
Medical EMI/EMC Compliance
All models include built-in EMI filtering to meet the
following emissions requirements:
EMI SPECIFICATIONS
Conducted Emissions
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Medical Safety
Condor D.C. Power Supplies, Inc. declares under our sole
responsibility that all GSM7 models are in conformity with
the applicable requirements of EN60601-1 following the
provisions of the Low Voltage Directive 73/23/EEC.
All GSM7 models are certified to be in compliance with
the applicable requirements of UL2601-1 (Patient Care
Equipment), CSA 22.2 No. 234 (with additional tests to
C22.2 No 601.1 per T.I.L. CA-08), EN 60601-1. Classification
(in accordance with sub-clause 5 of EN60601-1):
(5.1) Protection against electric shock = class II
(5.2) Degree of protection against electric shock = signal
output or intermediate
(5.3) Protection against harmful ingress of water =
ordinary (no protection)
(5.5) Not suitable for use in the presence of flammable
anesthetic mixture
(5.6) Mode of operation = continuous
Earth Leakage Current
Leakage current measured in ground wire connection to
output common.
* Normal condition **Single fault condition
FEATURES:
• World’s smallest medically approved 7 W supply
• 2.00" x 2.28" x 0.56"
• Wide-range ac input (85-264 Vac)
• Medical Approval to UL2601-1 and EN60601-1
• EMI conducted to CISPR11 Class B
• Ideal for ultra-compact applications
M
G
G
d o
S
S
M
M
l e
7
7
marked to LVD
L
a e
a k
2 3
5 5
e g
µ
µ
C
A
A
u
r r
n e
* t
L
a e
a k
e g
0 5
5 8
EN61000-4-5, 1 kV diff., 2 kV com.
EN55011 Class B; FCC Class B
EN61000-4-2, 6 kV contact, 8 kV air
EN61000-4-4, 2 kV, 5 kHz
COMPLIANCE LEVEL
EN61000-4-3, 3 V/meter
C
µ
µ
A
A
u
r r
n e
* * t
1
2
e T
4 6
2 3
t s
V
V
a
a
o V
/ c
/ c
a t l
0 6
0 5
e g
H
H
z
z
e T
U
E I
t s
L
C
6 2
M
6
e
1 0
1 0
h t
1 -
1 -
d o

Related parts for GSM7-15

GSM7-15 Summary of contents

Page 1

... GSM7 models are in conformity with the applicable requirements of EN60601-1 following the provisions of the Low Voltage Directive 73/23/EEC. All GSM7 models are certified compliance with the applicable requirements of UL2601-1 (Patient Care Equipment), CSA 22.2 No. 234 (with additional tests to C22.2 No 601.1 per T.I.L. CA-08), EN 60601-1. Classification (in accordance with sub-clause 5 of EN60601-1): (5 ...

Page 2

... GSM7 Medical 7 Watt Single Output GSM7 MECHANICAL SPECIFICATIONS AC NEUTRAL º Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033 800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com º Units should be allowed to warm up/operate under non-condensing conditions before application of power. B. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3 ...

Related keywords