80PK-3A Fluke Electronics, 80PK-3A Datasheet

no-image

80PK-3A

Manufacturer Part Number
80PK-3A
Description
PROBE SURFACE
Manufacturer
Fluke Electronics
Series
SureGrip™r
Type
K-Type Thermocoupler
Datasheets

Specifications of 80PK-3A

Usage
HVAC
Tip - Type
Contact, Spring
Plug Type
Sub-Mini (Thermocouple Plug)
Probe Temperature Range
32 ~ 500°F (0 ~ 260°C)
Cable Length
39.370" (1000.00mm)
Cable Insulation
Poly-Vinyl Chloride (PVC)
Plug Color
Yellow
Probe Length
3.750" (95.25mm)
Probe Material
PTFE (Polytetrafluoroethylene)
Lead Free Status / RoHS Status
Lead free by exemption / RoHS compliant by exemption
Other names
874610
TO AVOID ELECTRICAL SHOCK, DO NOT USE THIS PROBE
WHEN VOLTAGES EXCEEDING 24V AC RMS OR 60 V DC ARE
PRESENT.THE PROBE TIP IS ELECTRONICALLY CONNECTED
TO THE OUTPUT TERMINALS.
The thermocouple strip in the tip of the probe is designed to
deflect only .030” in normal operation. Any action which bends
or pulls the strip out further from the tip will CONSIDERABLY
reduce the life of the probe
INTRODUCTION
The 80PK-3A Type K Thermocouple Surface Probe is designed for
measuring the temperature of flat or slightly convex surfaces, with
an exposed junction to allow direct contact with the surface being
measured. The 40-inch (1-meter) cable is terminated with a Type K
miniature thermocouple connector with 0.792-mm (0.312 in) pin
spacing. The probe can be used with any temperature-measuring
instrument that is designed to accept type K thermocouples and has
a miniature connector input.
SPECIFICATIONS
Type: K (Chromel vs Alumel)
Measurement Range: 0 C to 260 C (32 F to 500 F)
Junction Accuracy: (With respect to ANSI MC96.1): ±2.2 C
(3.96 F) over the range of 0 C to 260 C (32 F to 500 F)
Restrictions: The 260 C continuous temperature rating is primarily
determined by the Teflon support piece. The Teflon insulation
should not be exposed to temperatures exceeding 260 C (500 F)
nor to open flame, since this can cause release of toxic material.
Output: 25 C (77 F) =1.00 mV (reference junction at 0 C)
Seebeck Coefficient: 25 C (77 F) = 40.50 V / C
Measurement Time: (Time Constant): 3 sec typical on metal
surface; 15 sec max. for a 260 C change. See Applications
Information below.
Maximum Voltage: 24V ac rms or 60V dc
Probe Tip:
September 1989 Rev. 4, 4/02
©1993-2002 Fluke Corporation. All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies
Type K Thermocouple Surface Probe
Maximum Temperature: 260 C (500 F)
Material: White PTFE
WARNING
CAUTION
Instruction Sheet
80PK-3A
®

80PK-3A Summary of contents

Page 1

... CONSIDERABLY reduce the life of the probe INTRODUCTION The 80PK-3A Type K Thermocouple Surface Probe is designed for measuring the temperature of flat or slightly convex surfaces, with an exposed junction to allow direct contact with the surface being measured ...

Page 2

... Protection: Class 3. Relates solely to insulation and grounding properties defined in IEC 348. MEASUREMENT CONSIDERATIONS Instrument Compatibility The 80PK-3A is designed to be compatible with any temperature- measuring instrument that accepts type K thermocouples, has a miniature thermocouple connector, and has cold reference junction compensation. Accuracy of the temperature measuring instrument could be considered along with the 80PK-3A accuracy specification in order to determine the overall accuracy of the combination ...

Page 3

MEASURING TECHNIQUE Here are some suggestions for improving the accuracy of your temperature measurements: When measuring higher than ambient temperatures, adjust the connection between the probe and the surface until you get the highest temperature reading. When measuring lower than ...

Page 4

Fluke Corporation (Fluke) warrants this product to be free from defects in material and workmanship under normal use and service for a period of 1 year from date of shipment. Software is warranted to operate in accordance with its programmed ...