M210 RHOPOINT, M210 Datasheet - Page 2

Resistance Meter

M210

Manufacturer Part Number
M210
Description
Resistance Meter
Manufacturer
RHOPOINT
Datasheet

Specifications of M210

Wire Connection
4-Wire
Test Current
5mA
External Height
15cm
External Width
80mm
External Depth
38mm
Weight
480g
Length/height, External
150mm
Operating Temperature Min
5°C
Operating Temperature Max
35°C

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
M210-A1A
Manufacturer:
IODMR
Quantity:
20 000
Part Number:
M2100-20U6999L
Manufacturer:
TE
Quantity:
5 400
Part Number:
M2100-20U6999L-208/240VAC
Manufacturer:
TE
Quantity:
5 400
Company:
Part Number:
M2101S020
Quantity:
500
Company:
Part Number:
M2101S020
Quantity:
500
Part Number:
M21036G-12
Manufacturer:
NNDSPEE
Quantity:
87
Part Number:
M21036G-12
Manufacturer:
MINDSPEED
Quantity:
20 000
Part Number:
M21038/27
Manufacturer:
AD
Quantity:
24
Part Number:
M21038/27-01 Q1553-1
Manufacturer:
a
Quantity:
2
Company:
Part Number:
M21038/27-02
Quantity:
580
Part Number:
M21038/27-13
Manufacturer:
PULSE
Quantity:
8 000
Company:
Part Number:
M21038/27-26
Quantity:
364
Description
This battery operated digital resistance meter is designed to meas-
ure accurately low resistance from 0.001 ohm to 200 ohms, using
four-terminal measurment network which eliminates errors caused
by test lead resistance.
in one hand with only an ON button to operate. It has auto-ranging,
auto-decimal point and auto-powerdown after three minutes. It has
The Measuring Circuit
Packing list
SET COMPRISING:
ONE Milli-ohmeter M210..............................................M210
ONE Set standard spring clip connectors and leads...........M210/9A
ONE Instructions leaflet
OPTIONAL EXTRAS:
ONE Ever-ready case
SET Probe and contact connectors...............................M210/9B
The instrument is small and light weight and is used easily
Specification
Ranges (selected automatically)
Increment
Accuracy
Zero offset
Test current
Test power
Temp. drift
Operating range
Display
Weight
Dimensions
Approximate weight (instrument only)
Battery
(instrument incase with batteries
Approximate dimensions: Instrument
Case ...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
leads M21019A)
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
long battery life and is powered by a single PP3 or MN 1604 bat-
tery
Some typical applications
Relay . switch and connector contacts . PC-plated through holes .
PC track resistance . Transformer and motor windings . Ring circuit
testing . Earth conductor continuity . Wire-wrapped and solder joints
. RFI sheilding and coil resistance.
In the four-terminal “Kelvin” configuration four wires are used to connect
the circuit under test to the instruments.
One pair of wires carries current generated by the instrument, the other
pair connect to the voltage-measuring circuit.
The current circuit is low impedance, the voltage circuit high impedance.
When connected a current flows through the component or circuit under
test and the voltage drop across the test points is measured.
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
...
1.999, 19.99, 199.9 Ohms
0.001, 0.01, 0.1 Ohms
±0.1% of range
±0.001 Ohms
5 to 35°C
510 gms.
H: 15cm, w: 8.Ocm, D: 3.8cm
H: 15.5cm, w: 14.5cm, D: 4.5cm
5mA Max
5mW Max
0.1%°C of lowest range
3½ digit L.C.D.
480 gms.
170 gms.
9v (PP3 MN 1604 or equiv.)
1EC6F22
Milli-ohneter being used to test a reed relay.
Note the ease of connection.