8808A 120V Fluke Electronics, 8808A 120V Datasheet - Page 11

DMM 5.5 RES RS-232 INTERFACE

8808A 120V

Manufacturer Part Number
8808A 120V
Description
DMM 5.5 RES RS-232 INTERFACE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8808A 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1068
Power Considerations
Selecting the Line Voltage
Replacing the Fuses
Line-Power Fuse
To store the Meter, place the box under cover in a location that complies with the storage
environment specifications described in the “General Specifications” section later in this
manual.
The Meter operates on varying power distribution standards found throughout the world
and must be set up to operate on the line voltage that will power it. The Meter is packed
ready for use with a line voltage determined at the time of ordering. If the selected line
voltage does not match the power that the Meter will be plugged into, the Meter’s line-
voltage setting must be changed and replacement of the line fuse may be required.
The Meter operates on four different input line voltages. The selected line-voltage setting
is visible through the window in the line-fuse holder on the Meter’s rear panel.
Changing the line-voltage setting may require a different line-power fuse for proper
operation.
The Meter uses one fuse to protect the line-power input and two fuses to protect current-
measurement inputs.
The Meter has a line-power fuse in series with the power supply. Table 3 indicates the
proper fuse for each of the four line-voltage selections. The line-power fuse is accessed
through the rear panel.
To avoid electric shock or fire, do not use makeshift fuses or
short-circuit the fuse holder.
1. Unplug the power cord.
2. Insert a small screwdriver blade into the narrow recess to the left of the fuse
3. Remove the voltage-selector block from the fuse holder.
4. Rotate the selector block until the desired voltage rating faces outward.
5. Replace the selector block back into the fuse holder.
6. Install the fuse holder back into the Meter and reconnect the power cord.
1. Unplug the power cord.
2. Insert a small screwdriver blade into the narrow recess to the left of the fuse
3. Remove the fuse and replace it with a fuse of an appropriate rating for the
4. Replace the selector block back into the fuse holder.
holder and pry it to the right until the holder pops out. See Figure 1.
holder and pry it to the right until the holder pops out. See Figure 1.
selected line-power voltage. See Table 3.
XW Warning
Power Considerations
Digital Multimeter