8808A 120V Fluke Electronics, 8808A 120V Datasheet - Page 25

DMM 5.5 RES RS-232 INTERFACE

8808A 120V

Manufacturer Part Number
8808A 120V
Description
DMM 5.5 RES RS-232 INTERFACE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8808A 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1068
AC Voltage Specifications
AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz,
add an additional error of 0.1 % of range, and for 50kHz to 100 kHz, add 0.13 % of range.
200 mV
2 V
20 V
200 V
1000 V
200 mV
2 V
20 V
200 V
750 V
200 mV
2 V
20 V
Range
Range
Notes:
[1]
Maximum Input ................................................. 750 V rms or 1000 V peak or 8 x 10
Measurement Method ....................................... AC-coupled true-rms. Measures the ac component of input with up to
AC Filter Bandwidth:
Common Mode Rejection................................. 60 dB at 50 Hz or 60 Hz (1 kΩ unbalance)
Maximum Crest Factor .................................... 3:1 at Full Scale
Additional Crest Factor Errors (<100 Hz) ...... Crest Factor 1-2, 0.05 % of full scale
Slow ............................................................... 20 Hz – 100 kHz
Range
Accuracy given as ± (% of reading + % of range)
(5-1/2 Digits)
50 kHz – 100 kHz
50 kHz – 100 kHz
50 kHz – 100 kHz
199.999 mV
20 kHz – 50 kHz
20 kHz – 50 kHz
20 kHz – 50 kHz
45 Hz – 20 kHz
45 Hz – 20 kHz
45 Hz – 20 kHz
Full-Scale
1.99999 V
19.9999 V
199.999 V
20 Hz – 45 Hz
20 Hz – 45Hz
20 Hz – 45Hz
750.00 V
Frequency
23 °C ± 5 °C
0.01 + 0.003
0.01 + 0.002
0.01 + 0.003
0.01 + 0.002
0.01 + 0.002
90 days
100 uV
10 mV
10 uV
Slow
1 mV
1 uV
23 °C ± 5 °C
Accuracy
0.15 + 0.05
0.15 + 0.05
0.15 + 0.05
0.8 + 0.05
0.3 + 0.05
0.8 + 0.05
0.8 + 0.05
0.3 + 0.05
0.8 + 0.05
0.8 + 0.05
0.3 + 0.05
0.8 + 0.05
90 days
Input Characteristics
1000 V dc bias on any range.
Crest Factor 2-3, 0.2 % of full scale
Accuracy
Accuracy
[1]
Resolution
0.015 + 0.004
0.015 + 0.003
0.015 + 0.004
0.015 + 0.003
0.015 + 0.003
Medium
1000 uV
100 mV
100 uV
23 °C ± 5°C
10 mV
10 uV
Accuracy
1 year
[1]
23 °C ± 5 °C
0.35 + 0.05
0.35 + 0.05
0.35 + 0.05
0.9 + 0.05
0.2 + 0.05
0.9 + 0.05
0.9 + 0.05
0.2 + 0.05
0.9 + 0.05
0.9 + 0.05
0.2 + 0.05
0.9 + 0.05
1 year
1000 uV
100 mV
100 uV
10 mV
10 uV
Fast
7
Temperature Coefficient/°C
Volts-Hertz product
Outside 18 – 28 °C
0.0015 + 0.0005
0.0020 + 0.0005
0.0015 + 0.0005
0.0015 + 0.0005
0.001 + 0.0005
1 MΩ ±2 % shunted by
Input Impedance
Outside 18 – 28 °C
Coefficient/°C
Temperature
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.05 + 0.01
0.05 + 0.01
0.05 + 0.01
<100 pf
Electrical Specifications
Digital Multimeter
3