8808A/SU 120V Fluke, 8808A/SU 120V Datasheet - Page 3

MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT

8808A/SU 120V

Manufacturer Part Number
8808A/SU 120V
Description
MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT
Manufacturer
Fluke
Datasheet

Specifications of 8808A/SU 120V

Dmm Type
Bench
Voltage Measuring Range Dc
200mV-1000V
Voltage Measuring Range Ac
200mV-750V
Current Measuring Range Dc
200µA To 10A
Current Measuring Range Ac
20mA To 10A
Lead Free Status / RoHS Status
na
2
Fluke Corporation
Accuracy specifications are valid for 5-½ digit mode and after at least a half-hour warm-up.
DC Voltage Specifications
8808A
200 mV
2 V
20 V
200 V
1000 V
200 mV
2 V
20 V
200 V
1000 V
Range
Notes:
[1]
Notes:
[1]
Maximum Input ........................................................1000 V on any range.
Common Mode Rejection .......................................120 dB at 50 or 60 Hz
Normal Mode Rejection ..........................................80 dB at Slow Rate
A/D Nonlinearity ......................................................15 ppm of range
Input Bias Current ...................................................<30 pA at 25 C
Settling Considerations..........................................Measurement settling times are affected by source impedance, cable
Range
Accuracy given as ± (% of reading + % of range)
At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M .
1.99999 V
199.999 V
1000.00 V
(5-1/2 Digits)
199.999 mV
19.9999 V
Full-Scale
0.01 + 0.003
0.01 + 0.002
0.01 + 0.003
0.01 + 0.002
0.01 + 0.002
23 °C ± 5 °C
90 days
100 µV
10 mV
10 µV
Slow
1 mV
1 µV
Accuracy
Input Characteristics
dielectric characteristics, and input signal changes
[1]
Electrical Specifications
Accuracy
Resolution
Fluke 8808A Digital Multimeter Extended Specifications
1000 µV
Medium
100 mV
100 µV
0.015 + 0.004
0.015 + 0.003
0.015 + 0.004
0.015 + 0.003
0.015 + 0.003
10 mV
10 µV
23 °C ± 5°C
1 year
0.1
1000 µV
100 mV
100 µV
10 mV
(1 k unbalance)
10 µV
Fast
Temperature Coefficient/°C
Outside 18 – 28 °C
0.0015 + 0.0005
0.0020 + 0.0005
0.0015 + 0.0005
0.0015 + 0.0005
0.001 + 0.0005
Input Impedance
10 M ±1 %
10 M ±1 %
10 M ±1 %
>10 G
>10 G
Fluke Corporation 3
[1]
[1]