8808A/SU 120V Fluke, 8808A/SU 120V Datasheet - Page 6

MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT

8808A/SU 120V

Manufacturer Part Number
8808A/SU 120V
Description
MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT
Manufacturer
Fluke
Datasheet

Specifications of 8808A/SU 120V

Dmm Type
Bench
Voltage Measuring Range Dc
200mV-1000V
Voltage Measuring Range Ac
200mV-750V
Current Measuring Range Dc
200µA To 10A
Current Measuring Range Ac
20mA To 10A
Lead Free Status / RoHS Status
na
6 Fluke Corporation
AC Current
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs
from 1 % to 5 % of range, add an additional error of 0.1 % of range.
200 µA
2 mA
20 mA
200 mA
2 A
10 A
20 mA
200 mA
2 A
10 A
20 mA
200 mA
2 A
10 A
Range
Notes:
[1]
Notes:
[1]
Input Protection .......................................................Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses
Measurement Method .............................................AC-coupled True RMS
Shunt Resistance .....................................................0.01 for 2 A and 10 A ranges
AC Filter Bandwidth:
Maximum Crest Factor............................................3:1 at Full Scale
Additional Crest Factor Errors (<100 Hz) .........Crest Factor 1-2, 0.05 % of full scale
Range
Range
Slow ........................................................................20 Hz – 100 kHz
Accuracy given as ± (% of reading + % of range)
Accuracy given as ± (% of reading + % of range)
(5-1/2 Digits)
45 Hz - 2 kHz
45 Hz - 2 kHz
45 Hz - 2 kHz
45 Hz - 2 kHz
20 Hz - 45Hz
20 Hz - 45Hz
20 Hz - 45Hz
20 Hz - 45Hz
19.9999 mA
199.999 mA
Frequency
Full-Scale
1.99999 A
10.0000 A
Fluke 8808A Digital Multimeter Extended Specifications
0.015 + 0.005
0.02 + 0.005
0.02 + 0.005
23 °C ± 5 °C
0.03 + 0.02
0.05 + 0.02
0.18 + 0.01
90 days
100 µA
0.1 µA
Accuracy
10 µA
Slow
1 µA
23 °C ± 5 °C
0.25 + 0.05
0.25 + 0.05
0.25 + 0.05
0.8 + 0.05
0.35 + 0.1
1 + 0.05
1 + 0.05
90 days
1 + 0.1
Input Characteristics
1 for 20 mA and 200 mA
Crest Factor 2-3, 0.2 % of full scale
[1]
Accuracy
Accuracy
0.03 + 0.005
0.02 + 0.005
0.03 + 0.008
23 °C ± 5 °C
0.04 + 0.02
0.08 + 0.02
0.2 + 0.01
1 year
Resolution
Accuracy
Medium
100 µA
10 µA
1 mA
1 µA
[1]
23 °C ± 5 °C
1.25 + 0.06
1.25 + 0.06
1.25 + 0.12
0.3 + 0.06
0.3 + 0.06
0.3 + 0.06
0.5 + 0.12
1 + 0. 06
1 year
100 µA
10 µA
1 mA
1 µA
Fast
Temperature Coefficient/°C
Outside 18 – 28 °C
0.003 + 0.001
0.002 + 0.001
0.005 + 0.001
0.005 + 0.001
0.008 + 0.001
0.008 + 0.001
Outside 18 – 28 °C
Burden Voltage
Coefficient/°C
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
Temperature
<0.05 V
<0.5 V
<0.1 V
<0.5 V
8808A Fluke Corporation 5