E4404B STD AGILENT TECHNOLOGIES, E4404B STD Datasheet - Page 14

ANALYZER, SPECTRUM, 9KHZ To 6.7GHZ

E4404B STD

Manufacturer Part Number
E4404B STD
Description
ANALYZER, SPECTRUM, 9KHZ To 6.7GHZ
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of E4404B STD

Spectrum Analyzer Type
Bench
Frequency Measuring Range
9kHz To 6.7GHz
Sensitivity Dbm
-139dBm
External Height
222mm
External Width
416mm
External Depth
409mm
Weight
37.7lb
Lead Free Status / RoHS Status
na
Temperature range
Data storage (nominal)
Memory usage (nominal)
Measurement speed
General Specifications
Weight (without options)
a.
b.
c.
Operating
Storage
Disk drive
EMI compatibility
Audible noise
sound pressure at 25 °C
Military specifications
Power requirements
AC operation on (line |)
Standby (line )
DC operation
Internal b
External
State
State plus 401- point trace
Kilograms
Local measurement rate
Remote measurement and
GPIB transfer
RF center freq tuning time
Display resolution
14
Meeting class A performance during DC operation.
For serial numbers < US414400 or MY41440000, 1MB without Option B72,
8 Mb with Option B72.
401 sweep points. The size of a state will increase depending on the installed
application(s).
d
Conducted and radiated interference is in compliance with CISPR Pub. 11/1990 Group 1 Class B a (Option 060)
E4411B
≥ 35/sec
≥ 30/sec
13.2 kg
29.1 lb
≤ 90 ms
Conducted and radiated interference is in compliance with CISPR Pub. 11/1990 Group 1 Class A
Type tested to the environmental specifications of MIL-PRF-28800F class 3
Type tested to the environmental specifications of MIL-PRF-28800F class 3
Basic analyzer
E4403B
≥ 30/sec
≥ 30/sec
≤ 90 ms
15.5 kg
34.2 lb
< 40 dBa pressure and < 4.6 bels power (ISODP7779)
200 traces or states/8.0 MB
12 to 20 Vdc, < 200 W power consumption
≤ 90 ms
E4408B
≥ 28/sec
≥ 30/sec
17.1 kg
37.7 lb
90 to 132 V rms, 47 to 440 Hz
195 to 250 V rms, 47 to 66 Hz
-40 °C to +75 °C
Power consumption < 300W
10 °C to +40 °C
0 °C to +55 °C
Power consumption < 5W
3.5” 1.44 MB, MS-DOS
d.
640 x 480
The LCD display is manufactured using high precision technology. However, there
may be up to six bright points (white, blue, red or green in color) that constantly
appear on the LCD screen. These points are normal in the manufacturing process
and do not affect the measurement integrity of the product in any way.
20 kB c
16 kB c
Standard, communications test analyzer
E4402B
≥ 45/sec
≥ 45/sec
15.5 kg
34.2 lb
≤ 75 ms
or custom configuration
E4404/05/07B
≥ 40/sec
≥ 40/sec
17.1 kg
37.7 lb
≤ 75 ms