DS1615 Maxim Integrated Products, DS1615 Datasheet - Page 20

IC RECORDER TEMP 5V 16-DIP

DS1615

Manufacturer Part Number
DS1615
Description
IC RECORDER TEMP 5V 16-DIP
Manufacturer
Maxim Integrated Products
Type
Temperature Recorderr
Datasheet

Specifications of DS1615

Memory Size
32B
Time Format
HH:MM:SS (12/24 hr)
Date Format
YY-MM-DD-dd
Interface
3-Wire Serial
Voltage - Supply
4 V ~ 5.5 V
Operating Temperature
-40°C ~ 85°C
Mounting Type
Through Hole
Package / Case
16-DIP (0.300", 7.62mm)
Lead Free Status / RoHS Status
Contains lead / RoHS non-compliant

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
DS1615S
Manufacturer:
MAXIM/美信
Quantity:
20 000
Note that the Specification test command is ignored if the ST button is pulled to ground when the
command is issued.
4. Read Temperature (55h)
Host Transmit:
DS1615 Response (Host Receives): When the device is not currently datalogging (i.e., MIP = 0), the
Temperature is immediately measured and value is written to the Current Temperature register.
This command instructs the DS1615 to immediately perform a temperature measurement and to store the
resulting value in the Current Temperature register. The temperature value obtained from this command
is not stored in the datalog or histogram memory. After this command has been executed, the user must
read the Temperature Ready (TR) bit in the Status register to determine if the temperature measurement
has been completed. If the TR bit is a logic 1, the measurement has been completed and the value in the
Current Temperature register is valid. If the TR bit is a logic 0, the measurement has not been completed.
This command functions only when MIP = 0 (i.e., the device is not currently datalogging). If MIP =1, the
DS1615 takes no action in response to the command.
5. Clear Memory (A5h)
Host Transmit:
DS1615 Response: The contents of the datalog, histogram, Temperature Alarm, Current Samples, Start
Time Stamp, Start Delay, and Sample Rate registers are cleared if the Clear Memory command has been
enabled by setting the CLR bit in the Control register to a one. After clearing the memory, the MEM
CLR bit in the Status register is set. The Clear Memory command functions only if the oscillator is
active. The DS1615 is inaccessible for 500 µ s after the Clear Memory command has been issued.
0
1
D7
D7
1
0
D6
D6
0
1
D5
D5
1
0
D4
D4
20 of 24
0
0
D3
D3
1
1
D2
D2
0
0
D1
D1
1
1
D0
D0

Related parts for DS1615