U1602B-001 Agilent Technologies Test Equipment, U1602B-001 Datasheet - Page 4

no-image

U1602B-001

Manufacturer Part Number
U1602B-001
Description
Digital Oscilloscope, Handheld, Orange, 20 MHz with USB Port
Manufacturer
Agilent Technologies Test Equipment
Type
Digitalr
Datasheet

Specifications of U1602B-001

Bandwidth
20 MHz
Brand/series
U1600 Series
Configuration
Handheld
Display Type
Color LCD
Memory
125 Kpts
Number Of Channels
2
Sample Rate
0.2 GS/s
SCOPE SPECIFICATIONS
SCOPE CHARACTERISTICS
[1] All specifications are warranted. Specifications are valid after a 30-minute warm-up period and within a range of ±10 ºC from
[2] All characteristics are typical performance values and are not warranted. Characteristics are valid after a 30-minute warm-up
[3] Maximum sampling rates are shown here. Sampling rate will vary according to selected time base. For more information, please refer
[4] 200 MS/s sampling rate is only available at 125 ns/div timebase
[5] Number of points displayed will vary according to time base selected. For more information, refer to User’s and Service Guide
Vertical System: Scope Channels
Bandwidth (–3 dB)
DC vertical gain accuracy
Scope Channel Triggering
Trigger sensitivity
Acquisition: Scope Channels
Maximum sample rate
Equivalent sample rate
Vertical resolution
Maximum recording length:
Peak detection
Average
Vertical System: Scope Channels
Analog channels
Bandwidth (–3 dB)
AC coupled
Rise time
Single shot bandwidth
Vertical sensitivity
Maximum input
Offset/Dynamic range
Input impedance
Coupling
firmware calibration temperature.
period and within a range of ±10 ºC from firmware calibration temperature.
to User’s and Service Guide.
[1]
U1602A/U1602B: DC to 20 MHz
U1604A/U1604B: DC to 40 MHz
5 mV/div to 20 mV/div: ± 5% full scale
50 mV/div to 100 V/div: ± 3% full scale
DC to 5 MHz: 0.8 divisions
U1602A/U1602B: 5 MHz to 20 MHz – 1 division
U1604A/U1604B: 5 MHz to 40 MHz – 1 division
100 MSa/s per channel (50 s/div to 250 ns/div)
200 MSa/s single channel and interleaved (125 ns/div)
U1604A/U1604B: 2.5 GSa/s (125 ns/div to 10 ns/div)
8 bits
U1602A/U1604A: Up to 11 100 points, viewable on screen with zoom function
U1602B/U1604B: Up to 125 000 points, viewable on screen with zoom function
5 ns
Selectable in average number of 2, 4, 8, 16, 32, 64, 128, 256
Channel 1 and Channel 2 simultaneous acquisition
U1602A/U1602B: DC to 20 MHz
U1604A/U1604B: DC to 40 MHz
< 10 Hz without probe
< 1 Hz with 10 MW 10:1 probe
U1602A/U1602B: < 17.5 ns
U1604A/U1604B: < 8.8 ns
U1602A/U1602B: 20 MHz
U1604A/U1604B: 40 MHz
5 mV/div to 100 V/div (1:1 scope probe)
50 mV/div to 1 kV/div (10:1 scope probe)
500 mV/div to 10 kV/div (100:1 scope probe)
CAT III 300 Vrms (up to 400 Hz) from terminal to ground
± 5 div
1 MW I I < 20 pF
AC, DC, GND
[2]
4
3
4