DSO6032A-GSA AGILENT TECHNOLOGIES, DSO6032A-GSA Datasheet - Page 21

69R7191

DSO6032A-GSA

Manufacturer Part Number
DSO6032A-GSA
Description
69R7191
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of DSO6032A-GSA

Scope Bandwidth
300 MHz
Scope Type
Digital Bench
Scope Channels
2 Scope
Bandwidth
300MHz
Meter Display Type
XGA LCD
Sampling Rate
2GSPS
Input Impedance
1Mohm
Rise Time
1.17ns
Input Voltage
400V P-p
Rohs Compliant
NA
Performance characteristics
Trigger system (continued)
Nth edge burst
Autoscale
Scope channel triggering
Range (internal)
Sensitivity*
Coupling
Digital (D15 - D0) channel triggering (MSO6000A or MSO-upgraded DSO6000A only)
Threshold range (user defined)
Threshold accuracy
Predefined thresholds
External (EXT) triggering
Input impedance
Maximum input
Range
Sensitivity
Coupling
Probe ID
*
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
Trigger on the Nth edge of a burst that occurs after an idle time that you specify. Max edge
count: 65,536.
Finds and displays all active scope and logic (for MSO6000A series MSO) channels, sets edge
trigger mode on highest-numbered channel, sets vertical sensitivity on scope channels and
thresholds on logic channels, time base to display ~1.8 periods. Requires minimum voltage
>10 mVpp, 0.5% duty cycle and minimum frequency >50 Hz.
±6 div from center screen
<10 mV/div: greater of 1 div or 5 mV; ≥10 mV/div: 0.6 div
AC (~3.5 Hz on MSO/DSO601xA, ~10 Hz on MSO/DSO603xA/605xA/610xA),
DC, noise reject, HF reject and LF reject (~50 kHz)
±8.0 V in 10 mV increments
±(100 mV + 3% of threshold setting)
TTL = 1.4 V, CMOS = 2.5 V, ECL = –1.3 V
MSO/DSO6xx2A (2-/2+16-ch models)
MSO/DSO6012A: 1 MΩ ± 3% || 11 pF or 50 Ω
MSO/DSO6032A/6052A/6102A:
CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk
With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50-Ω input
DC coupling: trigger level ±1 V and ±8 V
For ±1 V range setting: DC to 100 MHz, 100 mV;
For ±8 V range setting: DC to 100 MHz, 250 mV;
AC (~3.5 Hz), DC, noise reject, HF reject and LF reject (~50 kHz)
MSO/DSO601xA: Auto probe sense
MSO/DSO603xA/605xA/610xA: Auto probe sense and AutoProbe interface
Agilent- and Tektronix-compatible passive probe sense
1 MΩ ± 3% || 14 pF or 50 Ω
MSO/DSO6032A/6052A/6102A:
>100 MHz to bandwidth of oscilloscope: 200 mV
MSO/DSO6032A/6052A/6102A:
>100 MHz to bandwidth of oscilloscope: 500 mV
(continued)
MSO/DSO6xx4A (4-/4+16-ch models)
MSO/DSO6014A: 1.015 kΩ ±5%
MSO/DSO6034A/6054A/6104A:
±15 V
±5 V
MSO/DSO6014A:
MSO/DSO6034A/6054A/6104A:
2.14 kΩ ±5%
DC to 100 MHz: 500 mV
DC to 500 MHz: 500 mV
21