MD1812 Supertex, Inc., MD1812 Datasheet
MD1812
Related parts for MD1812
MD1812 Summary of contents
Page 1
... DMOS FETs for medical ultrasound applications but may be used in any application that needs a high output current for a capacitive load. The input stage of the MD1812 is a high-speed level translator that is able to operate from logic input signals of 1.2 to 5.0V amplitude. An adaptive threshold circuit is used to set the level translator threshold to the average of the input logic 0 and logic 1 levels ...
Page 2
... quiescent current NEGQ NEG I V average current average current average current NEG NEG 16-Lead QFN 4x4mm body, MD1812K6-G Pin Configuration Value -0.5V to +13.5V V -0. +0. -0. +0. -7.0V to +0.5V V -10V to V +0. -0.5V to GND +7.0V SS +125°C -65° ...
Page 3
DC Electrical Characteristics ( 12V GND = 0V NEG Sym Parameter V Input logic voltage high IH V Input logic voltage low IL I Input logic current high ...
Page 4
... VH. All inputs must be kept low until the device is powered up. Good trace practices should be followed corresponding to the desired operating speed. The internal circuitry of the MD1812 is capable of operating up to 100MHz, with the primary speed limitation being the loading effects of the load capacitance. ...
Page 5
Logic Truth Table Logic Inputs OE INA INC Pin Description Pin # Function Description 1 INB Logic input. Controls ...
Page 6
... Dimensions marked with (*) are non-JEDEC dimensions. The value listed is for reference only. Drawings not to scale. (The package drawing(s) in this data sheet may not reflect the most current specifications. For the latest package outline information go to http://www.supertex.com/packaging.html.) Doc.# DSFP-MD1812 A051408 θ ...