EDS10-11C FUJI [Fuji Electric], EDS10-11C Datasheet

no-image

EDS10-11C

Manufacturer Part Number
EDS10-11C
Description
FC SERIES MOVING COIL TYPE INDICATOR
Manufacturer
FUJI [Fuji Electric]
Datasheet
such as FC series signal, DC voltage and current signals as
well as thermocouple and resistance bulb signals.
FEATURES
1. Various types of signals can be input directly to the in-
2. High input impedance
3. The indicator is compact and light weight.
SPECIFICATIONS
Measuring input: FC series signal
DATA SHEET
This is an indicator compatible with various types of inputs
The instrument has an input impedance of 1 M for input
of 1 to 5V DC. No circuit adjusting resistor is required for
thermocouple input.
dicator.
FC SERIES
DC current
DC voltage
Thermocouple F; 0 to 10mV DC or more
Thermocouple G;10mV DC span or
Resistance bulb H;
Resistance bulb W;
A; 1 to 5V DC
B; 4 to 20mA DC
D; 0 to 100 A
E; 0 to 10mV ... 5V DC
(With reference junc-
tion compensator)
(Upper limit overshoot
burnout circuit is avail-
able on request)
more
(With reference junc-
tion compensator)
(Upper limit overshoot
burnout circuit is avail-
able on request)
JPt100
Pt100
MOVING COIL TYPE INDICATOR
50 C span or more,
three-wire type
50 C span or more,
three-wire type
...200mA DC
Input resistance and allowable input signal source resistance;
Input signal
Thermocouple
Resistance
Slide resistance (resistance span 100 )
Opening angle transmitter L:
1 to 5V
Others
4 to 20mA
Others
J: Three wire type
K: Two-wire type
Three-wire type, resistance
span 100 (Zero point and span
adjustable)
Input
resistance
1M
40k to 1M
5
1 to 100
40 to 600k
(with circuit adjusting resis-
tor)
Allowable input
signal source
resistance
2k or less
100 or less
100 or less
(50 or less
when burnout cir-
cuit is provided)
6 or less per wire
Date
EDS10-11c
PBA
Fab. 28, 1991

Related parts for EDS10-11C

EDS10-11C Summary of contents

Page 1

... Allowable input Input signal source resistance resistance less 40k to 1M 100 or less 5 — 100 — 100 or less (50 or less 40 to 600k when burnout cir- cuit is provided) — less per wire EDS10-11c Date Fab. 28, 1991 ...

Page 2

PBA Allowance: 1.0% Scale length: 100mm Response time:Within 3 sec (time required for stabilizing the indicator pointer within 1.5% after applying input corresponding to 2/3 (approx.) of full scale) Power supply: 24V (20 to 30V 100V 10% AC, ...

Page 3

OUTLINE DIAGRAM (Unit : mm) CONNECTION DIAGRAM Resistance Resistor bulb 24V DC (or Power supply 100V AC) Voltage Input Current Alarm ...

Page 4

PBA ORDERING INFORMATION 1. Product name 2. Code symbol 3. Input 4. Scale 5. Whether or not alarm device and alarm indicator lamp are required 6. Power supply 7. Application 8. Other necessary information Caution on Safety *Before using this ...

Related keywords