DPS53-M

Manufacturer Part NumberDPS53-M
DescriptionAC/DC Power Supply Single-OUT 12V 5A 60W 3-Pin
ManufacturerAstec Power
DPS53-M datasheet
 

Specifications of DPS53-M

Input Voltage90 to 264 VOutput Voltage12 V
Maximum Current Rating2(rms) AMaximum Output Current5 A
Number Of Outputs1Maximum Output Power60 W
Efficiency80(Typ) %Input Frequency47 to 440 Hz
Operating Temperature0 to 40 °C  
1
Page 1
2
Page 2
Page 1/2

Download datasheet (286Kb)Embed
Next
DPS50-M
60 Watts
Medical
Total Power:
30 - 60 Watts
Input Voltage:
90 - 264 Vac
127 - 300 Vdc
# of Outputs:
Single
Special Features
• Medical safety approvals
• CE Mark EMC & LVD (except
DPS52-M)
• Universal AC input
• Fully regulated output
• Overload and short circuit
protection
• Internal remote sense to the
output plug
• Constant voltage
• Level V efficiency (except
DPS52-M)
• High MTBF
• IEC320 AC receptacle, 3 pin
(type C14)
• Built in EMI filter
(CISPR 22 Class B)
• LED power good indicator
• Input power <74 watts
• Complies with EN61000-3-2
• Meets Energy Star
• WEEE & CeC iV compliant
Safety
UL
UL 60601-1
CSA
CSA-C22.2 no.601.1-M90
VDE
07501/EN60601-1 (IEC601)
EMC & LVD (except
CE
DPS52-M)
Electrical Specifications
Input
90 - 264 Vac (wide range)
Input ra n ge
127 - 300 Vdc
Frequency
47 - 440 Hz
Inrush current
<60 A @ 230 Vac, cold start @ 25 ˚C
Input current
<2.0 A rms @ 115 Vac; <1A @ 230 Vac
Efficiency
80% @ full load typical, @115 Vac
EMI/RFI
FCC Part 15, Subpart B Class B & EN55022 (CISPR 22) Class B
Safety ground
275 µA max. @ 50/60 Hz, 264 Vac input
leakage current
Output
Maximum Power (Po)
60 W (DPS52, 30 W)
10ms. typical at full load @ 115 Vac
Hold-up time
Overvoltage protection
20% - 45% above nominal output maximum
Latching type, recycle AC to reset.
Output short circuit protection auto recover
Overload protection
Overload protected @ 120 - 150% above maximum rating
Cable/connector
DC cable with 2.5 mm center plug
DC plug center +v
DC plug outer -v
Environmental Specifications
Operating temperature:
0° to +40 °C ambient. Derate each output as 2.5% per degree from
40°C to 60 °C. -20°C start up
Storage temperature:
-40 °C to +85 °C
Electromagnetic
Designed to meet EN61000-4, -2, -3, -4, -5, -6, -8, -11, level 3
susceptibility:
Humidity:
Operating; non-condensing 10% to 90% RH
Three orthogonal axes, 3 sweeps per axe at 1 oct/min. 2G peak 8Hz
Vibration:
to 500 Hz, operational
MTBF demonstrated:
> 550,000 hours at full load and 25 °C ambient conditions
Embedded Power for
Business-Critical Continuity
Rev. 03.16.11_123
DPS50-M
1 of 2

DPS53-M Summary of contents

  • Page 1

    DPS50-M 60 Watts Medical Total Power Watts Input Voltage 264 Vac 127 - 300 Vdc # of Outputs: Single Special Features • Medical safety approvals • CE Mark EMC & LVD (except DPS52-M) • Universal AC input • Fully regulated output • Overload and short circuit protection • Internal remote sense to the output plug • Constant voltage • Level V efficiency (except DPS52-M) • High MTBF • IEC320 AC receptacle, 3 pin (type C14) • Built in EMI filter (CISPR 22 Class B) • LED power good indicator • Input power <74 watts • Complies with EN61000-3-2 • Meets Energy Star ...

  • Page 2

    ... Ordering Information Model Maximum Output Maximum Number Power Voltage Load DPS52- DPS53 DPS54 DPS55 2.5 A DPS58 1. Peak current lasting <15 seconds with a maximum 10% duty cycle ºC including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings. 3. Peak-to-peak with 20 MHz bandwidth and 10 µF (tantalum capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges approval under new EUP directive Mechanical Drawing 5.24" (133.0) AC INLET MODEL LABEL Notes: 1. Specifications subject to change without notice. ...