EL-USB-4 Lascar Electronics, EL-USB-4 Datasheet - Page 2

no-image

EL-USB-4

Manufacturer Part Number
EL-USB-4
Description
Data Logger, Current (4-20mA) with Direct USB Interface
Manufacturer
Lascar Electronics
Type
Loggerr
Datasheet

Specifications of EL-USB-4

Accuracy
±1 %
Range, Measurement
4 to 20 mA
Resolution
0.05 mA
Sample Rate
Every 1 Second to Every 12 Hours
Temperature, Operating, Range
–25 (–13) to 80 (176) °C (°F)
Lead Free Status / Rohs Status
RoHS Compliant part
EL-USB-4
Specifications liable to change without prior warning
- Green single flash,
- Green single flash,
- Red single flash,
- Red double flash
- Green single flash,
- Red single flash,
- Red double flash,
- Green double flash,
- Red/Green single flash,
- No LEDs flash
LED FLASHING MODES
EL-USB-4 features a red and a green LED.
The LEDs on an EL-USB-4 will flash in one of the following ways, assuming alarm hold is turned off (exact timings to
be confirmed).
By default hold is disabled. In this mode the red LED will no longer continue to flash after the logged temperature
has returned to normal from an alarm condition.
Hold can be turned on via the control software. In this mode the red LED that indicates an alarm condition will
continue to flash, even after the logged temperature has returned to normal. The red LED will effectively have
latched into its alarm condition. This feature ensures that the user is notified that an alarm level has been exceeded,
without the need to download the data from the logger.
every 30 seconds
every 10 seconds
every 10 seconds
every 10 seconds
every 20 seconds
every 20 seconds
every 20 seconds
every 20 seconds
every 20 seconds,
alarm hold turned on
Logger primed and ready to start.
Logging in progress, last stored value OK
Logging in progress, the last stored reading is equal to or exceeds the low alarm level
(If hold is turned on, then a stored value is equal to or exceeds the low alarm level)
Logging in progress, the last stored reading is equal to or exceeds the high alarm level
(If hold is turned on, then a stored value is equal to or exceeds the high alarm level)
Low battery, logging in progress, the last stored value OK
Low battery, logging in progress, the last stored value is equal to or exceeds the low alarm level
Low battery, logging in progress, the last stored value is equal to or exceeds the high alarm level
Logger is full, last stored value OK
Logger is full, last stored value is equal to or exceeds the low or high alarm level
LEDs disabled during configuration to save battery power and increase logging duration.
or Battery dead.
www.lascarelectronics.com
EL-USB-4
4-20mA Current Loop USB Data Logger
Issue 4
04/2008
S.C.
No: 0754 - Page 2 of 3
Applies to EL-USB-4/1