EL-USB-4 Lascar Electronics, EL-USB-4 Datasheet - Page 3

no-image

EL-USB-4

Manufacturer Part Number
EL-USB-4
Description
Data Logger, Current (4-20mA) with Direct USB Interface
Manufacturer
Lascar Electronics
Type
Loggerr
Datasheet

Specifications of EL-USB-4

Accuracy
±1 %
Range, Measurement
4 to 20 mA
Resolution
0.05 mA
Sample Rate
Every 1 Second to Every 12 Hours
Temperature, Operating, Range
–25 (–13) to 80 (176) °C (°F)
Lead Free Status / Rohs Status
RoHS Compliant part
EL-USB-4
Specifications liable to change without prior warning
CONNECTION
BATTERY REPLACEMENT
APPLICATIONS
We recommend that you replace the battery every 12 months, or prior to logging critical data.
The EL-USB-4 does not lose its stored readings when the battery is discharged or when the battery is replaced; the
data logging process will however be stopped and cannot be re-started until the battery has been replaced and the
logged data has been downloaded to PC.
Only use 3.6V 1/2AA lithium batteries. Check with your supplier that the battery you are ordering is 'press fit' and
is not fitted with solder tags. Before replacing the battery, remove the EL-USB-4 from the PC.
Note:
Leaving the EL-USB-4 plugged into the USB port for longer than necessary will cause some of the battery capacity
to be lost.
current loop
transmitter
4-20mA
WARNING: Handle lithium batteries carefully, observe warnings on battery casing.
EL-USB-4 fitted to a 4-20mA
current loop transmitter
Dispose of in accordance with local regulations.
other current
loop devices
www.lascarelectronics.com
EL-USB-4
4-20mA Current Loop USB Data Logger
Issue 4
04/2008
4-20mA probe head
EL-USB-4 fitted to a
Positive measurement input
Negative measurement input
S.C.
No: 0754 - Page 3 of 3
Applies to EL-USB-4/1

Related parts for EL-USB-4