72-6851 TENMA, 72-6851 Datasheet - Page 28

POWER SUPPLY, BENCH, 35V

72-6851

Manufacturer Part Number
72-6851
Description
POWER SUPPLY, BENCH, 35V
Manufacturer
TENMA
Datasheet

Specifications of 72-6851

Power Supply Output Type
Variable
No. Of Outputs
1
Output Voltage
35V
Output Current
10A
Load Regulation
0.01%
Length
350.52mm
Width
210.82mm
Height
132.08mm
Output Type
Variable
Calibrated
No
Rohs Compliant
NA
Lead Free Status / RoHS Status
na
DELTAV?
DELTAI?
INCVV
INCV
DECVV
DECV
INCI
Tenma 72-6851 & 72-6853 Instruction manual
Sequential command
Operation complete message generated immediately after <rmt> is sent.
Returns the delta voltage value in Volts in <nr2> numeric format. The syntax of the
response is
Example: If the delta voltage is 0·55 Volts the response to the command
Sequential command
Operation complete message generated immediately after <rmt> is sent.
Returns the delta current value in Amps in <nr2> numeric format. The syntax of the
response is
Example: If the delta current is 0·550 Amps the response to the command
Sequential command
Operation complete message generated immediately after execution.
Increment the output voltage by the delta voltage value and verify that the voltage is within
±3 counts or 5% of the target value. If the value of the output voltage is outside the range
for the specified output the value is set to the maximum value allowed, no error is
generated.
Sequential command
Operation complete message generated immediately after execution.
This command is identical to INCVV above except that no verification of the output voltage
is performed, thus saving the possible 500ms required to read back the output voltage.
This is useful when it is known that the slew time will be short or during constant current
operation.
Sequential command
Operation complete message generated immediately after execution.
Decrement the output voltage by the delta voltage value and verify that the voltage is
within ±3 counts or 5% of the target value. If the value of the output voltage is outside the
range for the specified output the value is set to the minimum value allowed, no error is
generated.
Sequential command
Operation complete message generated immediately after execution.
This command is identical to DECVV above except that no verification of the output
voltage is performed, thus saving the possible 500ms required to read back the output
voltage. This is useful when it is known that the slew time will be short or during constant
current operation.
Sequential command
Operation complete message generated immediately after execution.
Increment the set current by the delta current value. If the value of the set current is
outside the range for the specified output the value is set to the maximum value allowed,
no error is generated.
www.tenma.com
DELTAV? will be DELTAV 0.55<rmt>
DELTAI? will be DELTAI 0.550<rmt>
DELTAV <nr2><rmt>
DELTAI <nr2><rmt>
27