34410A AGILENT TECHNOLOGIES, 34410A Datasheet - Page 15

MULTIMETER, DIGITAL, BENCH, 6-1/2 DIGIT

34410A

Manufacturer Part Number
34410A
Description
MULTIMETER, DIGITAL, BENCH, 6-1/2 DIGIT
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 34410A

Dmm Type
Bench
Voltage Measuring Range Dc
100mV To 1000V
Voltage Measuring Range Ac
100mV To 750V
Current Measuring Range Dc
100µA To 3A
Current Measuring Range Ac
100µA To 3A
Lead Free Status / RoHS Status
na
The aperture setting is set to
MIN. For the 34410A, this
would be 100 µs. For the
34411A, 20 µs is the MIN. The
same commands can be used
for DCI, ohms, or temperature
– but you have to change the
VOLT to CURR, RES, FRES,
or TEMP.
Autozero cannot be turned off
when making four-wire ohms
measurements. Temperature
measurements are all resistive
and can be two- or four-wire
measurements.
200 Hz - 300 kHz (Fast)
20 Hz - 300 kHz (Medium)
3 Hz - 300 kHz (Slow)
Input Frequency
2.5 seconds/reading
0.625 second/reading
0.025 second/reading
Default Settling Delay
How to set up the DMM
for fast AC measurements
For AC measurements, the
following are the key factors
in measurement speed:
• AC filter setting
• Autorange
• Automatic trigger delay
All AC measurements are limited
to 500 readings/s or less.
Measurements can be slower
dependent upon the settings of
these three factors. AC filter is
common to all AC measurements
and is the most significant
factor. The response of the
filter affects measurement
speed. See the table below:
Maximum Reading Rate
167 readings/second
500 readings/second
50 readings/second
15
The default setting when using
the automatic trigger delay is
shown. If the trigger delay is
set to zero, the maximum
reading rate is possible.
Autorange does not significantly
impact the measurement speed
until a high-voltage signal causes
an actuator relay to change
ranges. The automatic trigger
delay slows down the measure-
ment speed based mainly upon
the selection of the AC filter.
SCPI command example to
achieve maximum measurement
speed for ACV:
CONF:VOLT:AC
VOLT:AC:BAND 200
VOLT:AC:RANGE 1.0
TRIG:DELAY:AUTO OFF
TRIG:DELAY 0
ACI will use the same com-
mands – except for specifying
CURR instead of VOLT. For
frequency, the FREQ:APER
command specifies a gate time
of 0.001, 0.01, 0.1, or 1 second,
which will also affect reading
rates in addition to the AC
filter selection.
Peak-to-peak measurements
do not slow down AC RMS
measurements.