53230A Agilent Technologies Test Equipment, 53230A Datasheet - Page 13

no-image

53230A

Manufacturer Part Number
53230A
Description
Universal Frequency Counter/Timer, 350 MHz, 12 digit/sec, 20 ps
Manufacturer
Agilent Technologies Test Equipment
Type
Counter/Timerr
Datasheet

Specifications of 53230A

Frequency Range
DC coupled 1 mHz to 350 MHz; AC coupled 10 Hz to 350 MHz
Input Impedance
Selectable 1 MΩ ±1.5% or 50Ω ±1.5%
Input Type
Front panel BNC
Input, Range
±5 V (± 50 V) full scale ranges
Noise
500 μVrms (max), 350 μVrms (typ)
Number Of Channels
3 Channels
Sensitivity
DC - 100 MHz: 20 mVpk > 100 MHz: 40 mVpk
General Characteristics (nom)
continued
53210A
53220A
53230A
Battery (Option 300)
Technology
Internal lithium ion battery with integrated smart battery monitor & charger
Operating temperature
0 to 55º C Battery will only charge under 35º C
limits
Instrument running on battery power above 50º C
will turn off to minimize battery capacity degradation.
Storage temperature
-10º C to 60º C.
limits
Extended exposure to temperatures above 45° C could degrade battery
performance and life
Operating time (typ)
3 hours when operated below +35º C
Standby time - OCXO
24 hours
powered (typ)
Recharge time (typ)
1
4 hours to 100% capacity; 2-hours to 90% capacity
Accessories included
CD
User's guide, SCPI/programmers reference, programming examples, drivers (IVI-COM,
LabView), IO library instructions
Cables
Power line cord, 2 m USB 2.0
Warranty
Standard
1-year
1. Assumes calibrated battery.
212.8 mm
261.2 mm
103.8 mm
88.3 mm
272.3 mm
302.2 mm
Dimensions apply to all three models: 53210A, 53220A, 53230A.