53230A Agilent Technologies Test Equipment, 53230A Datasheet - Page 6

no-image

53230A

Manufacturer Part Number
53230A
Description
Universal Frequency Counter/Timer, 350 MHz, 12 digit/sec, 20 ps
Manufacturer
Agilent Technologies Test Equipment
Type
Counter/Timerr
Datasheet

Specifications of 53230A

Frequency Range
DC coupled 1 mHz to 350 MHz; AC coupled 10 Hz to 350 MHz
Input Impedance
Selectable 1 MΩ ±1.5% or 50Ω ±1.5%
Input Type
Front panel BNC
Input, Range
±5 V (± 50 V) full scale ranges
Noise
500 μVrms (max), 350 μVrms (typ)
Number Of Channels
3 Channels
Sensitivity
DC - 100 MHz: 20 mVpk > 100 MHz: 40 mVpk
Measurement Characteristics
Time interval A to B, B to A
Range
Time interval A or B
Range
Minimum width
Minimum edge repetition
rate
Level & slope
Single-period, pulse-width, rise time, fall time
Range
Minimum width
Minimum edge repetition
Rate
Level & slope
Duty
Range
Minumim width
Level & slope
Phase A to B, B to A
Range
Totalize measurements
Channels
Range
Rate
Gating
Level measurements
Voltage level - standard
input channels
Microwave power level
(microwave channel option)
9
6
9
53210A
±5.1 Vpk with 2.5 mV resolution or ±51 Vpk with 25 mV resolution
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
continued
6
0 to 4 relative signal power
2 ns (assumes at least 4 ns from stop to start edge)
2 ns (assumes at least 4 ns from stop to start edge)
(assumes at least 4 ns from stop to start edge)
.000001 to .999999 or 0.0001% to 99.9999%
Continuous, timed, or external gate input
53220A
Pulse width, rise/fall time: 2 ns
Auto-level or user selectable
Auto-level or user selectable
Auto-level or user selectable
-0.5 ns to 100,000 s (min)
-1 ns to 100,000 s (nom)
2 ns to 100,000 s (min)
Gate accuracy is 20 ns
-180.000º to 360.000º
Single-period: 2.5 ns
0 to 10
0 s to 1000 s
Ch 1 or Ch 2
0 - 350 MHz
6 ns
6 ns
15
events
53230A