53230A Agilent Technologies Test Equipment, 53230A Datasheet - Page 4

no-image

53230A

Manufacturer Part Number
53230A
Description
Universal Frequency Counter/Timer, 350 MHz, 12 digit/sec, 20 ps
Manufacturer
Agilent Technologies Test Equipment
Type
Counter/Timerr
Datasheet

Specifications of 53230A

Frequency Range
DC coupled 1 mHz to 350 MHz; AC coupled 10 Hz to 350 MHz
Input Impedance
Selectable 1 MΩ ±1.5% or 50Ω ±1.5%
Input Type
Front panel BNC
Input, Range
±5 V (± 50 V) full scale ranges
Noise
500 μVrms (max), 350 μVrms (typ)
Number Of Channels
3 Channels
Sensitivity
DC - 100 MHz: 20 mVpk > 100 MHz: 40 mVpk
Input Channel Characteristics
1. AC coupling occurs after 50 ohm termination
2. When ordered with optional rear terminals, the standard/baseband channel inputs are active on both the front and rear of the
3. Multiply value(s) by x10 for the 50 V range.
4. Stated specification assumes Noise Reject OFF. Noise Reject ON doubles the sensitivity.
5. Assumes sine wave.
Optional Microwave Inputs (nom)
Frequency range
Option 106
Option 115
Input
Connector
Input impedance (typ)
Input coupling
Amplitude range
Option 106
Option 115
Sensitivity (typ)
Input event thresholds
Level range
AM tolerance (CW only)
Maximum input
Damage level
universal counter though the specifications provided only apply to the rear terminals. Performance for the front terminals with rear
terminals installed is not specified.
5
53210A
Option 203 moves the input connector to a rear panel SMA(f)
continued
Auto-ranged for optimum sensitivity and bandwidth
Autoranged to +15 dBm max. (1.25 Vrms)
Autoranged to +19 dBm max. (2 Vrms)
6 GHz (Opt 106): -27 dBm (10 mVrms)
Front panel precision Type-N(f)
4
2 GHz – 13 GHz: -26 dBm
13 GHz – 15 GHz: -21 dBm
0.3 GHz – 2 GHz: -23 dBm
50 Ω ± 1.5% (SWR < 2.5)
50% modulation depth
> +27 dBm (5 Vrms)
300 MHz - 15 GHz
15 GHz (Opt 115):
100 MHz - 6 GHz
53220A
AC
53230A