N9010A-526 Agilent Technologies Test Equipment, N9010A-526 Datasheet - Page 4

no-image

N9010A-526

Manufacturer Part Number
N9010A-526
Description
EXA SIGNAL ANALYZER 26 GHZ
Manufacturer
Agilent Technologies Test Equipment
Type
Networkr
Datasheet

Specifications of N9010A-526

Frequency
26 GHz
Lead Free Status / Rohs Status
RoHS Exempt Product
Frequency and Time Specifications
Frequency range
Option 503
Option 507
Option 513
Option 526
Band
0
1
1
2
3
4
Frequency reference
Accuracy
Aging rate
Temperature stability
Achievable initial calibration accuracy
Example frequency reference accuracy
(with Option PFR) 1 year after
last adjustment
Residual FM
Frequency readout accuracy (start, stop, center, marker)
± (marker frequency x frequency reference accuracy + 0.25% x span + 5% x RBW + 2 Hz + 0.5 x horizontal resolution
1. Horizontal resolution is span/(sweep points – 1)
Marker frequency counter
Accuracy
Delta counter accuracy
Counter resolution
4
20 to 30 °C
5 to 50 °C
Option PFR
Standard
LO Multiple (N)
1
1
1
2
2
4
DC Coupled
9 kHz to 3.6 GHz
9 kHz to 7.0 GHz
9 kHz to 13.6 GHz
9 kHz to 26.5 GHz
9 kHz to 3.6 GHz
3.5 to 7.0 GHz
3.5 to 8.4 GHz
6.9 to 13.6 GHz
13.5 to 17.1 GHz
17 to 26.5 GHz
± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy]
Option PFR
±1 x 10
±1.5 x 10
Option PFR
±1.5 x 10
±5 x 10
Option PFR
±4 x 10
= ±(1 x 1 x 10
= ±1.9 x 10
≤ (0.25 Hz x N) p-p in 20 ms nominal
≤ (10 Hz x N) p-p in 20 ms nominal
See band table above for N (LO Multiple)
± (marker frequency x frequency reference accuracy + 0.100 Hz)
± (delta frequency x frequency reference accuracy + 0.141 Hz)
0.001 Hz
–7
–8
–8
–7
–8
/ year
–7
/ 2 years
–7
+ 5 x 10
–8
+ 4 x 10
–8
)
AC Coupled
10 MHz to 3.6 GHz
10 MHz to 7.0 GHz
10 MHz to 13.6 GHz
10 MHz to 26.5 GHz
Standard
±1 x 10
Standard
±2 x 10
±2 x 10
Standard
±1.4 x 10
Required options
Option 507
Options 513, 526
–6
–6
–6
–6
/ year
1
)