N9010A-526 Agilent Technologies Test Equipment, N9010A-526 Datasheet - Page 6

no-image

N9010A-526

Manufacturer Part Number
N9010A-526
Description
EXA SIGNAL ANALYZER 26 GHZ
Manufacturer
Agilent Technologies Test Equipment
Type
Networkr
Datasheet

Specifications of N9010A-526

Frequency
26 GHz
Lead Free Status / Rohs Status
RoHS Exempt Product
Frequency and Time Specifications
Analysis bandwidth
Maximum bandwidth
1. Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for further analysis or processing
Video bandwidth (VBW)
Range
Accuracy
Measurement speed
Local measurement and display
update rate
Remote measurement and LAN
transfer rate
Marker peak search
Center frequency tune and transfer (RF)
Center frequency tune and transfer (µW) 47 ms nominal
Measurement/mode switching
2. Sweep points = 101
6
in the time, frequency, or modulation domain.
1
2
Option B25 25 MHz
Standard
1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz and wide open (labeled 50 MHz)
±6% nominal
Option PC2
4 ms (250/s) nominal
5 ms (200/s) nominal
1.5 ms nominal
20 ms nominal
39 ms nominal
10 MHz
(continued)
Standard
11 ms (90/s) nominal
6 ms (167/s) nominal
5 ms nominal
22 ms nominal
49 ms nominal