Description
Specifications: Scope Type: Bench ; Scope Channels: 4 Analogue + 16 Digital + 1 RF ; Bandwidth: 500MHz ; Meter Display Type: XGA TFT-LCD Color ; Sampling Rate: 2.5GSPS ; Input Impedance: 1Mohm ; Rise Time: 700ps ; Input Voltage: 300Vrms ; Supply Volt
Manufacturer
Tektronix
Datasheet
Display Characteristics
Input/Output Ports
LAN eXtensions for Instrumentation (LXI)
Power Source
Characteristic
Display Type
Display Resolution
Waveform Styles
Graticules
Format
Waveform Capture Rate
Port
USB 2.0 High-speed Host
Port
USB 2.0 Device Port
LAN Port
XGA Video Port
Probe Compensator
Output
Auxiliary Out
External Reference In
Kensington Lock
VESA Mount
Characteristic
Class
Version
Characteristic
Power Source Voltage
Power Source Frequency 45 to 66 Hz (85 to 264 V)
Power Consumption
Description
10.4 in. (264 mm) liquid-crystal TFT color display
1,024 horizontal × 768 vertical pixels (XGA)
Vectors, Dots, Variable Persistence, Infinite
Persistence
Full, Grid, Solid, Cross Hair, Frame, IRE, and mV
YT and simultaneous XY/YT
>50,000 wfm/s maximum
Description
Supports USB mass storage devices, printers and
keyboard. Two ports on front and two ports on rear of
instrument
Rear-panel connector allows for communication/control
of oscilloscope through USBTMC or GPIB
(with a TEK-USB-488), and direct printing to all
PictBridge-compatible printers
RJ-45 connector, supports 10/100/1000 Mb/s
DB-15 female connector, connect to show the
oscilloscope display on an external monitor or projector
Front-panel pins
Amplitude: 2.5 V
Frequency: 1 kHz
Rear-panel BNC connector
V
V
to ground
Output can be configured to provide a pulse out signal
when the oscilloscope triggers, the internal oscilloscope
reference clock out, or an event out for limit/mask
testing
Time-base system can phase lock to an external
10 MHz reference (10 MHz ±1%)
Rear-panel security slot connects to standard
Kensington lock
Standard (MIS-D 100) 100 mm VESA mounting points
on rear of instrument
Description
LXI Class C
V1.3
Description
100 to 240 V ±10%
360 to 440 Hz (100 to 132 V)
225 W maximum
OUT
OUT
(Hi): ≥2.5 V open circuit, ≥1.0 V 50 Ω to ground
(Lo): ≤0.7 V into a load of ≤4 mA; ≤0.25 V 50 Ω
Physical Characteristics
Environmental
Ordering Information
MDO4000 Family
All Models Include: One passive voltage probe per analog channel (TPP0500
500 MHz, 10X, 3.9 pF for 500 MHz models; TPP1000 1 GHz, 10X, 3.9 pF for
1 GHz models), P6616 16-channel Logic Probe, Logic Probe Accessory Kit
(020-2662-xx), Front Cover (200-5130-xx), N-to-BNC Adapter (103-0045-00), User
Manual (071-2918-xx), Documentation CD (063-4367-xx), OpenChoice
Software, NI LabVIEW SignalExpress™ Tektronix Edition Software, Calibration
Certificates document measurement traceability to National Metrology Institute(s) –
the Quality System this product is manufactured in is ISO9001 registered, power
cord, accessory bag (016-2030-xx) and a three-year warranty. Please specify power
plug and manual language version when ordering.
Dimensions
Height
Width
Depth
Weight
Net
Shipping
Rackmount Configuration 5U
Cooling Clearance
Characteristic
Temperature
Humidity
Altitude
Regulatory
Model
MDO4054-3
MDO4054-6
MDO4104-3
MDO4104-6
Operating
Nonoperating
Operating
Nonoperating
Operating
Nonoperating
Electromagnetic
Compatibility
Safety
Mixed Domain Oscilloscopes — MDO4000 Series
Description
0 ºC to +50 ºC
–20 ºC to +60 ºC
High: 40 ºC to 50 ºC, 10% to 60% Relative Humidity
Low: 0 ºC to 40 ºC, 10% to 90% Relative Humidity
High: 40 ºC to 60 ºC, 5% to 60% Relative Humidity
Low: 0 ºC to 40 ºC, 5% to 90% Relative Humidity
9,843 ft. (3,000 m)
30,000 ft. (9,144 m)
EC Council Directive 2004/108/EC
UL61010-1, Second Edition; CSA61010-1 Second
Edition, EN61010-1:2001; IEC 61010-1:2001
Description
Mixed Domain Oscilloscope with (4) 500 MHz analog
channels, (16) digital channels, and (1) 3 GHz RF input
Mixed Domain Oscilloscope with (4) 500 MHz analog
channels, (16) digital channels, and (1) 6 GHz RF input
Mixed Domain Oscilloscope with (4) 1 GHz analog
channels, (16) digital channels, and (1) 3 GHz RF input
Mixed Domain Oscilloscope with (4) 1 GHz analog
channels, (16) digital channels, and (1) 6 GHz RF input
2 in. (51 mm) required on left side and rear of
instrument
mm
10.7
229
439
147
kg
5
www.tektronix.com
17.3
23.6
in.
9.0
5.8
lb.
11
®
Desktop