MDO4054-3GSA Tektronix, MDO4054-3GSA Datasheet - Page 7

no-image

MDO4054-3GSA

Manufacturer Part Number
MDO4054-3GSA
Description
Specifications: Scope Type: Bench ; Scope Channels: 4 Analogue + 16 Digital + 1 RF ; Bandwidth: 500MHz ; Meter Display Type: XGA TFT-LCD Color ; Sampling Rate: 2.5GSPS ; Input Impedance: 1Mohm ; Rise Time: 700ps ; Input Voltage: 300Vrms ; Supply Volt
Manufacturer
Tektronix
Datasheet
Normal, Average, Max Hold, and Min Hold spectrum traces.
Automated Channel Power measurement.
Spectrum Traces
The MDO4000 Series offers four different traces or views of the RF input
including Normal, Average, Max Hold, and Min Hold. You can set the
The optional TPA-N-VPI adapter enables any active, 50 Ω TekVPI probe to be connected
to the RF input.
detection method used for each trace type independently or you can leave
the oscilloscope in the default Auto mode that sets the detection type
optimally for the current configuration. Detection types include +Peak,
–Peak, Average, and Sample.
RF Measurements
The MDO4000 Series includes three automated RF measurements –
Channel Power, Adjacent Channel Power Ratio, and Occupied Bandwidth.
When one of these RF measurements is activated, the oscilloscope
automatically turns on the Average spectrum trace and sets the detection
method to Average for optimal measurement results.
RF Probing
Signal input methods on spectrum analyzers are typically limited to cabled
connections or antennas. But with the optional TPA-N-VPI adapter, any
active, 50 Ω TekVPI probe can be used with the RF input on the MDO4000
Series. This enables additional flexibility when hunting for noise sources
and enables easier spectral analysis by using true signal browsing on an
RF input.
Mixed Domain Oscilloscopes — MDO4000 Series
www.tektronix.com
7

Related parts for MDO4054-3GSA