EDS11-164A FUJI [Fuji Electric], EDS11-164A Datasheet - Page 2

no-image

EDS11-164A

Manufacturer Part Number
EDS11-164A
Description
Manufacturer
FUJI [Fuji Electric]
Datasheet
2
5. Operation and display section
6. Alarm (option)
7. Other functions
8. Power failure processing
9. Self-check
10. Operation and storage conditions
SOCKET
Parameter setting
method
Display
Status display LED
Setting accuracy
Indication accuracy
(at 23°C)
Alarm kind
Alarm ON-delay
Process alarm
output
Parameter mask
function
Ramp/soak
function (option)
Memory protection Held by non-volatile memory
Method
Ambient operating
temperature
Ambient operating
humidity
Storage temperature
What is alarm with hold?
The alarm is not turned ON immediately even when the process
value is in the alarm band. It turns ON when it goes out the alarm
band and enters again.
PXR4
Lower limit alarm
Lower limit alarm
(with hold)
Power ON
Digital setting by 3 keys
With key lock function
Process value/set value Independent display
4 digits, 7-segment LED
Control output, process alarm output
0.1% or less of measuring range
Thermocouple: ±(0.5% of measuring range)
For thermocouple R at 0 to 500°C ...
For thermocouple B at 0 to 400°C ...
Resistance bulb, voltage/current:
Absolute alarm, deviation alarm, zone alarm
with upper and lower limits for each
Hold function available (See the figure below.)
Alarm latch, Excitation/non-excitation
selecting function provided
Delay setting 0 to 9999 sec settable in 1 sec
steps
Relay contact: SPST contact: 220 V AC/30 V DC,
Mechanical life 10 million operations (no load)
Electrical life 100,000 operations (rated load)
Minimum switching current 100 mA (5 V DC)
MAX 2 points output cycle 0.5 sec
Parameter display is disabled by software.
2 program pattern of 4 steps each, or 1
program pattern × 8 steps
Digital input allows to start/reset the action.
Program error supervision by watchdog timer
-10°C to 50°C
( In low-temperature environment, start-up
time may vary in power activation.)
Less than 90% RH (no condensation)
-20°C to 60°C
± (1% of measuring range) ±1 digit ±1°C
± (5% of measuring range) ±1 digit ±1°C
± (0.5% of measuring range) ±1 digit
PV (process value)
Power OFF
off
off
on
±1 digit ±1°C
Power ON
Period where lower
limit alarm is output
off
on
on
1 A (resistive load)
off
off
Note 1: For current input connect the supplied 250Ω resister at
Note 2: Setting cannot be changed to a different group.
Note 3: When the measuring range exceeds 1000°C (1832°F),
11. Structure
Table 1 Measuring range table
Group
Mounting method
External terminal
Case material
Dimensions
Weight
Protective
structure
Outer casing
II
I
Resistance bulb
Thermocouple
DC voltage
DC current
the input terminal.
decimal point cannot be used.
input signal
Pt100
J
K
R
B
S
T
E
N
PL2
1 to 5V
4 to 20mA
Panel flush mounting, DIN rail mounting.
( Mounting socket is required for mounting
DIN rail. )
8 pins or 11 pins terminals
(Socket is required for wiring separately.)
Plastic
(non-combustible grade UL94V-0 equivalent)
48 × 48 × 84.7mm
Approx. 200 g
Front waterproof structure: NEMA4X
(IEC standard IP66 equivalent)
(when mounted on panel with our genuine
packing. Waterproof feature unavailable
in close mounting of multiple units)
Rear case: IEC IP20
Black (front frame, case)
measuring range(°C) measuring range(°F)
-150 to 850
-150 to 400
-150 to 800
scaling range
0 to 800
0 to 1200
0 to 1600
0 to 1800
0 to 1600
0 to 1300
0 to 1300
-238 to 1562
-238 to 752
-238 to 1472
-1999 to 9999
32 to 1472
32 to 2192
32 to 2912
32 to 3272
32 to 2912
32 to 2372
32 to 2372

Related parts for EDS11-164A