CDB5530U Cirrus Logic Inc, CDB5530U Datasheet - Page 11

no-image

CDB5530U

Manufacturer Part Number
CDB5530U
Description
BOARD EVAL FOR CS5530
Manufacturer
Cirrus Logic Inc
Type
A/Dr
Datasheets

Specifications of CDB5530U

Number Of Adc's
1
Number Of Bits
24
Sampling Rate (per Second)
3.84k
Data Interface
Serial
Inputs Per Adc
1 Differential
Input Range
±2.5 V
Power (typ) @ Conditions
35mW @ 5 V
Voltage Supply Source
Analog and Digital, Dual ±
Operating Temperature
-40°C ~ 85°C
Utilized Ic / Part
CS5530
Product
Data Conversion Development Tools
Resolution
24 bit
Maximum Clock Frequency
4 MHz
Interface Type
USB
Supply Voltage (max)
5 V
Supply Voltage (min)
3.3 V
For Use With/related Products
CS5530
Lead Free Status / RoHS Status
Contains lead / RoHS non-compliant
Lead Free Status / RoHS Status
Lead free / RoHS Compliant, Contains lead / RoHS non-compliant
Other names
598-1158
are displayed in the Conversion Data box, as well as the Channel (always “1” for CS5530) and Overflow
indicator bits for each conversion. The software will halt the data collection process when the user presses
the Stop button.
Configuration Register: In the Configuration Register box, the contents of the Configuration register are
displayed, and can be modified by typing a hexadecimal value in the Hexadecimal: box, or by changing
any of the values below the Hexadecimal: box to the desired settings.
Note: When changing the value of the reset system bit to “1” (RS, bit 29 in the Configuration register), the part will
be reset, and all registers will return to their default values. It is a good idea to select the Update Registers button
after performing a reset to update the screen with the new register values.
Offset / Gain Registers: In the Offset Registers / Gain Registers box, the contents of the offset and gain
registers are displayed in hexadecimal. These registers can all be modified in the Calibration window,
which is accessed when the user selects the Calibration Window button.
Calibration Window: The Calibration window (Figure 6) permits the user to access the calibration regis-
ters of the CS5530 and also perform system offset or gain calibrations. Prior to performing calibrations
however, the Configuration register should be modified to configure the part for the desired operating
mode (see
should be performed before gain calibration to ensure accurate results. Refer to the CS5530 data sheet
for more details.
System Offset Calibrations:
System Gain Calibrations:
DS742DB1
1) Attach an input signal to the AIN± inputs that you want to calibrate as zero. Note that the grounded
2) Activate the system offset calibration by pressing the System OFFSET button.
3) When the calibration is complete, the screen will automatically update the offset register with the
1) Attach a full-scale input to the AIN± inputs of the CDB5530U.
2) Activate the system gain calibration by pressing the System GAIN button.
3) When the calibration is complete, the screen will automatically update the gain register with the
input must be within the common mode range of the amplifier for proper operation.
new value.
new values.
Section 2.2.4, "Configuration Window" on page 13
Figure 6. CDB5530U Calibration Window
for more details). Note that offset calibration
CDB5530U
11