5.7.8
The trigger point is an event to output a trigger when the specified address has been accessed. Up to four trigger
points can be set by using the trigger outputs (four channels) on the bus monitoring circuit of the emulator.
The settings of the trigger point are displayed and modified on the [Trigger] sheet.
Selecting [Add...] or the event point and [Edit...] from the popup menu in this window displays the [Set Address
For Trigger] dialog box.
This dialog box allows the user to specify the address to be accessed as the trigger output condition during the
user program execution. Enable or disable the trigger output point by checking the check box on the left in the
screen.
[Trigger1]:
[Trigger2]:
[Trigger3]:
[Trigger4]:
[Address]:
72
Setting Trigger Points
Enables the output of trigger channel 1.
Enables the output of trigger channel 2.
Enables the output of trigger channel 3.
Enables the output of trigger channel 4.
Sets the address condition of the channel.
Figure 5.28 [Set Address For Trigger] Dialog Box
Figure 5.27 [Event] Window ([Trigger] Sheet)