8432180000 Weidmuller, 8432180000 Datasheet - Page 43

no-image

8432180000

Manufacturer Part Number
8432180000
Description
WTS4 PT100/2,V,0-10V
Manufacturer
Weidmuller
Datasheet

Specifications of 8432180000

Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Screw connection
Tension clamp connection
Input
Input current
Input frequency
Accuracy
Measuring principle
Connection type
Pass through diameter
Measurement circuit voltage
Max. measuring circuit
Output
Current/voltage selectable
Output voltage
Offset voltage
Load resistance
Output signal limit
Output current
Offset current
Load resistance
Status LED
Temperature coefficient
Response time (10 ... 90%)
(safe separation)
Rated voltage
Surge voltage
Overvoltage category
Contamination class
Clearance and creepage distance
Test voltage
Supply
Supply voltage
Power consumption at rated voltage
Reverse polarity protection
Operating temperature range
Storage/transport
Factory setting
Dimensions L/H/B (mm)
Weight
Approvals
Dimensions and accessories see
* T
198
WAVE
Approvals:
Input / Output
Block diagram
Ordering data
Technical data*
Coordination of insulation acc. to DIN EN 50178, 04/98
General
Current Monitoring
U
input and output range
adjustable via DIP switch
no calibration required
true TRMS value measurements
hall sensor measurement
method
= 23 °C single module
CONTROL
configurable
WAS2 CMA
WAZ2 CMA
5/10A ac/dc
selectable with
analogue output
0 ... 20 mA/
4 ... 20 mA/
0 ... 10 V
WAS2 CMA 5/10A uc
WAZ2 CMA 5/10A uc
5 A uc/10 A uc selectable (without additional adjustment)
0 Hz - 2 kHz (True RMS to DC Converter)
1% (0 Hz - 1 kHz) Crest factor 3 FSR
2% (0 Hz - 2 kHz) Crest factor 5 FSR
Contact-free current monitoring using Hall sensor
Push-through connection
8 mm
400 Vac, higher voltages dependent on wire insulation
dependent on wire cross-section
0 ... 10 V
0 ... 20 mA
4 ... 20 mA
0 ... 10 V
max. 0.08 V
≥1 KΩ
approx. 13 V and. 24 mA
0/4 ... 20 mA
max.150 µA
≤ 600 Ω
green LED
ON–> OK; blinks –> signal out of range; OFF –> Error
≤ 650 ppm/K
typ. 700 ms
300 V
6 KV
lll
2
≥ 5.5 mm
4 kV eff
21.6 Vdc...24 Vdc...26.4 Vdc
50 mA at I
yes
0 ... 50 ºC
-20 ... +70 ºC
0 ... 5A uc; 4 ... 20 mA
92.4/112.4/22.5
150 g
CE, cUL
Page 298 + 308
Type
out
= 20 mA
Cat. No.
8526610000
8526620000
Qty.
1
1
WAS2 CMA
WAZ2 CMA
20/25/30A ac/dc
selectable with
analogue output
0 ... 20 mA/
4 ... 20 mA/
0 ... 10 V
WAS2 CMA 20/25/30A uc
WAZ2 CMA 20/25/30A uc
20/25/30 A uc selectable (without additional adjustment)
0 Hz - 2 kHz (True RMS to DC Converter)
1% (0Hz - 1KHz) Crest factor 3 FSR
2% (0Hz - 2KHz) Crest factor 5 FSR
Contact-free current monitoring using Hall sensor
Push-through connection
8 mm
400 Vac, higher voltages dependent on wire insulation
dependent on wire cross-section
0 ... 10 V
0 ... 20 mA
4 ... 20 mA
0 ... 10 V
max. 0.08 V
≥1 KΩ
approx. 13 V and. 24 mA
0/4 ... 20 mA
max.150 µA
≤ 600 Ω
green LED
ON–> OK; blinks –> signal out of range; OFF –> Error
≤ 650 ppm/K
typ.700 ms
300 V
6 KV
lll
2
≥ 5.5 mm
4 kV eff
21.6 Vdc...24 Vdc...26.4 Vdc
50 mA at I
yes
0 ... 50 ºC
-20 ... +70 ºC
0 ... 25 A uc; 4 ... 20mA
92.4/112.4/22.5
150g
CE, cUL
Page 298 + 308
Type
out
= 20 mA
Cat. No.
8545830000
8545840000
Qty.
1
1

Related parts for 8432180000