LPT62-M

Manufacturer Part NumberLPT62-M
DescriptionPWR SUPPLY 60W 5/12/12V TRPL MED
ManufacturerEmerson Network Power
SeriesLPT60
LPT62-M datasheet
 

Specifications of LPT62-M

Voltage - Output5V, ±12VNumber Of Outputs3
Power (watts)60WApplicationsMedical
Power Supply TypeSwitching (Open Frame)Voltage - Input85 ~ 264VAC
Mounting TypeChassis Mount1st Output5 VDC @ 7A
2nd Output12 VDC @ 3A3rd Output-12 VDC @ 700mA
Size / Dimension5" L x 3" W x 1.69" H (127mm x 76.2mm x 42.9mm)Power (watts) - Rated60W
Operating Temperature0°C ~ 50°CEfficiency70%
ApprovalsCE, CSA, UL, VDELine Regulation±2%, ±5%
Load Regulation±2%, ±5%Output Power Rating35 W
Input Voltage85 VAC to 264 VACOutput Voltage (channel 1)5 V
Output Current (channel 1)7 AOutput Voltage (channel 2)12 V
Output Current (channel 2)3 AOutput Voltage (channel 3)- 12 V
Output Current (channel 3)0.7 AMounting StyleChassis
Dimensions5 in L x 3 in W x 1.65 in HLead Free Status / RoHS StatusLead free / RoHS Compliant
4th Output-  
1
Page 1
2
Page 2
3
Page 3
Page 1/3

Download datasheet (417Kb)Embed
Next
LPT60-M Series
60 Watts
Medical
Total Power:
60 - 80 Watts
Input Voltage:
85 - 264 VAC
120 - 300VDC
# of Outputs:
Triple
Special Features
• Medical safety approvals
• Universal input
• 3" x 5" footprint
• Remote sense on main output
• Built-in EMI filter
• Low output ripple
• Adjustable 5V output
• Overvoltage protection
• Overload protection
• 110KHz switching frequency
• LPX60 enclosure kit available
Safety
VDE
0750/EN60601-1 (IEC601)
21310-3336-0015
UL
UL2601
E182560
CSA
CSA 22.2 No. 601.1
LR109492
CE
Mark (LVD)
Electrical Specifications
Input
85-264VAC
Input ra nge
120-300VDC
Frequency
47-440Hz
Inrush current
<18A peak @ 115VAC;
<36A peak @ 230VAC, cold start @ 25°C
Input current
1.5A max. (RMS) @ 115VAC
Efficiency
70% typical at full load
EMI
FCC Class A conducted; CISPR 22 Class A conducted
EN55022 Class A conducted;
VDE 0878 PT3 Class A conducted
<75μA max. @ 50/60 Hz, 264VAC input
Safety ground
leakage current
Output
Maximum Power
60W for convection; 80W with 30CFW forced air
Cross regulation
±2% on output 1; ±5% on outputs 2, 3
Adjustment range
-5, +10% minimum
Hold-up time
20ms @ 60W load, 115VAC nominal line
Short circuit protection on all outputs. Case overload
Overload protection
protected @ 110% to 145% above peak rating
5.7 to 6.7 VDC on the main output
Overvoltage protection
Remote sense
Compensates for 0.5 V lead drop min. Will operate without
remote sense connected. Reverse connection protected.
Embedded Power for
Business-Critical Continuity
Rev. 06.15.07
LPT60-M Series
RoHS
1 of 3

LPT62-M Summary of contents

  • Page 1

    LPT60-M Series 60 Watts Medical Total Power Watts Input Voltage 264 VAC 120 - 300VDC # of Outputs: Triple Special Features • Medical safety approvals • Universal input • 3" x 5" footprint • Remote ...

  • Page 2

    ... Temperature coefficient: ±.04% per °C MTBF demonstrated: >550,000 hours at full load and 25°C ambient conditions Ordering Information Maximum Load Model Output Minimum with Convnection Number Voltage Load cooling LPT62-M +5V 0.7A 7A +12V 0.3A 3A -12V 0A 0.7A LPT63-M +5V 0.7A 7A +15V ...

  • Page 3

    ... Mechanical Drawing Pin Assignments Connector LPT62-M LPT63-M SK1-1 Neutral Neutral SK1-3 Line Line SK2-1 +12V +15V SK2- 2 +5V +5V SK2- 3 +5V +5V SK2-4 Common Common SK2-5 Common Common SK2-6 -12V -15V SK201-1 +Sense +Sense SK201-2 -Sense -Sense Mating Connectors AC Input: Molex 09-50-8031 (USA) ...