FLUKE-287/FVF Fluke Electronics, FLUKE-287/FVF Datasheet - Page 82

COMBO KIT DMM & ACCESSORY

FLUKE-287/FVF

Manufacturer Part Number
FLUKE-287/FVF
Description
COMBO KIT DMM & ACCESSORY
Manufacturer
Fluke Electronics
Series
FlukeView® Forms, 28xr
Type
Digital (DMM)r
Datasheets

Specifications of FLUKE-287/FVF

Includes
Battery, Case, Clips, Software, Test Leads
Style
Handheld
Display Digits
4.5
Display Type
LCD, Bar Graph
Display Count
50000
Function
Voltage, Current, Resistance, Capacitance, Temperature, Frequency
Functions, Extra
Continuity, dB
Features
Backlight, Data Logging, Hold, Memory, Min/Max/Ave
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
287/FVF
3340186
614-1151
287/289
Users Manual
Resistance Specifications
Temperature Specifications
72
Resistance
Conductance
[1]
[2]
[3]
[1]
[2]
When using the relative mode (REL Q) to compensate for offsets.
Add 20 counts above 33 nS in 50 nS range.
289 only.
Does not include error of the thermocouple probe.
Accuracy specification assumes ambient temperature stable to ±1 C. For ambient temperature changes of ±5 C, rated accuracy applies after 2 hours.
Function
-200 C to +1350 C
-328 F to +2462 F
Temperature
50
500
5 k
50 k
500 k
5 M
30 M
50 M
50 M up to 100 M
100 M up to 500 M
50 nS
[1]
[1][3]
[2]
[1]
[1]
Range
Resolution
0.1 C
0.1 F
Resolution
0.0001 M
0.0001 k
0.001 M
0.001 k
0.01 M
0.001
0.01 k
0.01 nS
0.1 M
0.1 M
0.01
Accuracy
1 % + 10
1 % + 18
0.15 % + 20
0.05 % + 10
0.05 % + 2
0.05 % + 2
0.05 % + 2
0.15 % + 4
Accuracy
1.5 % + 4
1.5 % + 4
3.0 % + 2
1 % + 10
8 % + 2
[1,2]