8846A/SU 120V Fluke Electronics, 8846A/SU 120V Datasheet - Page 37

PREC MLTMTR 6.5DIGI USB W/SFTWRE

8846A/SU 120V

Manufacturer Part Number
8846A/SU 120V
Description
PREC MLTMTR 6.5DIGI USB W/SFTWRE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8846A/SU 120V

Includes
Test Leads
Style
Bench
Display Digits
6.5
Display Type
VFD, Dual
Display Count
Varies
Function
Voltage, Current, Resistance, Capacitance, Temperature, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Memory, RS-232, USB Port
Ranging
Manual
Response
Average
Lead Free Status / RoHS Status
Contains lead / RoHS compliant by exemption
Other names
2675344
614-1003
Current-Input Fuses
The 400 mA and 10 A inputs are protected by user-replaceable fuses.
To test for a blown Current Input Fuse:
Line Voltage Selection
For protection against fire or arc flash, replace a blown fuse
only with one from Fluke.
The 400 mA input is protected by a fuse (F2) rated at 440 mA, 1000 V (fast blow),
10,000 A minimum breaking capacity (Fluke Part No. 943121).
The 10 A input is protected by a fuse (F1) rated at 11 A, 1000 V (fast blow),
10,000 A minimum breaking capacity (Fluke Part No. 803293).
1. With the Meter powered up, plug a test lead into the V GER connector.
2. Press N.
3. Insert the other end of the test lead into the 400 mA input connector.
4. Remove the probe from the 400 mA connector and insert into the 10 A
If the fuse is good, the Meter will read less than 200 . If the fuse is blown, the
Meter will read overload.
connector.
100
120
220
240
Table 2-1. Line Voltage to Fuse Rating
Figure 2-1. Replacing the Line Fuse
0.25 A, 250 V (slow blow)
0.25 A, 250 V (slow blow)
0.125 A, 250 V (slow blow)
0.125 A, 250 V (slow blow)
W Warning
Fuse Rating
Preparing the Meter for Operation
Fluke Part No.
166306
166306
166488
166488
Power Considerations
caw0201f.eps
2-5