Testing Continuity
To change the temperature scale, refer to the “Setting the Default Temperature Scale”
section in Chapter 3 of this manual. Available scales are Celsius, Fahrenheit, and Kelvin.
Refer to the “Adjusting the Meter’s Range” section in Chapter 3 of this manual for
information on how to adjust the measurement range.
Function modifiers:
Continuity testing determines whether a circuit is intact (i.e., has a resistance less than the
threshold). The threshold is selectable between 1 and 1000 .
To perform a continuity test:
1. Press S on the 8846A or Q on the 8845A. An example of the continuity display
Connect the test leads as shown in Figure 4-1.
To set the threshold value, refer to the “Setting the Continuity Threshold Resistance”
section in Chapter 3 of this manual.
Function modifiers:
The Beeper cannot be disabled in continuity. Pressing BEEPER OFF turns
off the beeper for errors.
4Wire
RTD 385
RO
ALPHA
2ND MEA
is shown below.
F1
F1
Switches the measurement input connectors to 4-wire measurement
for 4-wire RTDs. 4-wire RTDs make more accurate measurements.
Default RTD type. All coefficients are predefined.
Used to select a different value of RTD resistance at 0 C.
Used to set the first coefficient of the Calendar-Van Dusen equation.
- 246.645
Cycles the secondary display through the measurement functions
listed below and then off.
OHMS - Displays the resistance of the RTD. 2-wire resistance is used
when in 2-wire temperature mode and 4-wire resistance when the
Meter is in 4-wire RTD mode.
OPEN
F2
F2
F3
F3
Note
F4
F4
º
F
F5
F5
Making Measurements
Testing Continuity
caw11f.eps
caw12f.eps
4-15