8846A/SU 120V Fluke Electronics, 8846A/SU 120V Datasheet - Page 74

PREC MLTMTR 6.5DIGI USB W/SFTWRE

8846A/SU 120V

Manufacturer Part Number
8846A/SU 120V
Description
PREC MLTMTR 6.5DIGI USB W/SFTWRE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8846A/SU 120V

Includes
Test Leads
Style
Bench
Display Digits
6.5
Display Type
VFD, Dual
Display Count
Varies
Function
Voltage, Current, Resistance, Capacitance, Temperature, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Memory, RS-232, USB Port
Ranging
Manual
Response
Average
Lead Free Status / RoHS Status
Contains lead / RoHS compliant by exemption
Other names
2675344
614-1003
8845A/8846A
Users Manual
4-6
Function modifiers:
Filter
dB
dB€
2ND MEAS Cycles the secondary display through the measurement functions
3HZ SLOW Provides higher measurement accuracy on ac signals
20HZ
200HZ
Displays the filter menu. For best accuracy and stable readings,
choose a filter based on the lowest frequency to be measured and the
performance needed.
Displays the measured voltage as a decibel value referenced to a
stored relative value (dB = 20 log(Vnew/Vstored). The stored value is
obtained from the first measurement the Meter makes after pressing
the dB soft key. All future measurements are displayed using the
stored value as an offset. To take the Meter out of dB mode, press the
dB soft key.
Displays the measured voltage as a power decibel value referenced to
1 milliwatt (dBm = 10 log(power of Vnew into the reference
resistance / 1 mW) or 10 log(V
To accommodate the various impedances that a dBm measurement
can be made across, the Meter allows for the selection of 21 different
impedance values.
1. Press L.
2. Press the soft key labeled dB€ Ref.
3. With the impedance highlighted, press the soft key under the
listed below, and then off. When a second measurement function is
selected, the 2ND MEAS soft key label is highlighted.
VDC - Displays the dc voltage on which the ac signal may be riding.
Frequency - Displays the frequency of the ac signal applied to the
Input HI and LO connectors of the Meter.
ACV/DCV dual mode should not be used at frequencies
below 10 Hz.
To set the dB reference impedance:
value selected.
The available impedance settings are presented in sets of three
values. To move to a higher set of impedance values, press ++
-->. Press <-- -- to go to a lower set of impedance values.
between 3 Hz and 20 Hz. However, the measurement
cycle time is longer than it is when using the 20 Hz
filter.
Provides higher measurement accuracy on ac signals
between 20 Hz and 200 Hz. The measurement cycle
time is longer, however, than it is when using the
200 Hz filter.
Provides accurate measurements on ac signals 200 Hz
and above.
Note
2
/R*1 mW) where R is the resistance.