Specifications of N2735A OPTION 004

No. Of Channels
4
Scope Bandwidth
350 MHz
Scope Type
Hand Held
Lead Free Status / RoHS Status
na
18
Performance characteristics
Vertical system: scope channels (continued)
ESD tolerance
Noise, RMS, input shorted
DC vertical gain accuracy*
DC vertical offset accuracy
Single cursor accuracy
Dual cursor accuracy*
*
1
Vertical system: digital channels (MSO or MSO-upgraded DSO)
Number of channels
Threshold groupings
Threshold selections
User-defined threshold range
Maximum input voltage
Threshold accuracy*
Input dynamic range
Minimum input voltage swing
Input capacitance
Input resistance
Channel-to-channel skew
*
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
2 mV/div is a magnification of 4 mV/div setting for 350 MHz to 1 GHz models. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting and 32
mV for 2 mV/div sensitivity setting.
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
1
1
1
±2 kV
MSO/DSO701xB: 0.50% FS or 300 µV, whichever is greater
MSO/DSO703xB: 0.50% FS or 300 µV, whichever is greater
MSO/DSO705xB: 0.50% FS or 360 µV, whichever is greater
MSO/DSO710xB: 0.65% FS or 360 µV, whichever is greater
±2.0% full scale
≤200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value;
>200 mV/div: ±0.1 div ±2.0 mV ±1.5% offset value
±{DC vertical gain accuracy + DC vertical offset accuracy + 0.2% full scale (~1/2 LSB)}
Example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{2.0% (80 mV) + 0.1 (10 mV) + 2.0 mV + 0.5% (5 mV) + 0.2% (80 mV)} =
± 4.785 mV
±{DC vertical gain accuracy + 0.4% full scale (~1 LSB)}
Example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{2.0% (80 mV) + 0.4% (80 mV)} = ±1.92 mV
16 logic timing channels – labeled D15 - D0
Pod 1: D7 - D0
Pod 2: D15 - D8
TTL, CMOS, ECL and user-definable (selectable by pod)
±8.0 V in 10 mV increments
±40 V peak CAT I; transient overvoltage 800 Vpk
±(100 mV + 3% of threshold setting)
±10 V about threshold
500 mV peak-to-peak
~8 pF with flying leads
100 kΩ ±2% at probe tip
2 ns typical, 3 ns maximum
(continued)