Specifications of N2735A OPTION 004

No. Of Channels
4
Scope Bandwidth
350 MHz
Scope Type
Hand Held
Lead Free Status / RoHS Status
na
Performance characteristics
Trigger system (continued)
Selections
20
Edge
Pattern
Pulse width
TV
Sequence
CAN
FlexRay
LIN
USB
I
SPI
I
RS-232/UART
Duration
MIL-STD 1553
2
2
C
S
Edge, pulse width, pattern, TV, duration, sequence, CAN, LIN, USB, I
Nth edge burst, RS-232 with Option 232
Trigger on a rising, falling, alternating or either edge of any source
Trigger at the beginning of a pattern of high, low, and don’t care levels and/or a rising or
falling edge established across any of the analog and digital channels, but only after a
pattern has stabilized for a minimum of 2 nsec.
The scope channel’s high or low level is defined by that channel’s trigger level. The logic
channel’s trigger level is defined by the threshold for the pod, 0 - 7 or 8 - 15.
Trigger when a positive- or negative-going pulse is less than, greater than, or within a
specified range on any of the source channels.
Trigger using any scope channel on most analog progressive and interlaced video standards
including HDTV/EDTV, NTSC, PAL, PAL-M or SECAM broadcast standards. Select either
positive or negative sync pulse polarity. Modes supported include Field 1, Field 2, all fields,
all lines, or any line within a field. TV trigger sensitivity: 0.5 division of sync signal. Trigger
holdoff time can be adjusted in half field increments.
Arm on event A, trigger on event B (edge or pattern), with option to reset on event C or time delay.
Trigger on CAN (controller area network) version 2.0A and 2.0B signals. Trigger on the start
of frame (SOF) bit (standard). N5424A option supports triggering on remote frame ID (RTR),
data frame ID (~RTR), remote or data frame ID, data frame ID and data, error frame, all errors,
acknowledge error and overload frame.
Trigger on FlexRay Frames, errors, events and cycle-multiplexed triggering. N5432C or option
FLX supports also triggering on particular frame types symbolically, such as Startup frames,
Null frame, Sync frame, etc., as well as Boolean NOT frame types.
Trigger on LIN (local interconnect network) sync break at beginning of message frame
(standard). N5424A option supports triggering on frame ID.
USB Trigger on USB (universal serial bus) start of packet, end of packet, reset complete,
enter suspend, or exit suspend on the differential USB data lines. USB low speed and full
speed are supported.
Trigger on I
with address and/or data values. Also trigger on missing acknowledge, address with no acq,
restart, EEPROM read, and 10-bit write.
Trigger on SPI (serial protocol interface) data pattern during a specific framing period.
Supports positive and negative Chip Select framing as well as clock idle framing and
user-specified number of bits per frame.
This application provides triggering on audio bus protocol channels for audio left, right, either
as well as =, ≠, >,< entered data values and within and out of range values. It provides the
ability to easily view the audio packets on the waveform and in a listing window.
This application eliminates the need to manually decode bus traffic. Using data captured on
the scope or digital channels, the application provides the ability to easily view the information
sent over a RS-232 serial bus. Display real-time time-aligned decode of transmit and receive
lines. The application also enables triggering on RS-232/UART conditions.
Trigger on a multi-channel pattern whose time duration is less than a value, greater than a
value, greater than a time value with a timeout, or inside or outside of a set of time values.
Trigger on specific Command/Status Words, Data Words, and error conditions.
Minimum pulse width setting: 5 ns (MSO/DSO701xB/703xB scope channels)
Maximum pulse width setting: 10 s
Minimum duration setting: 2 ns
Maximum duration setting: 10 s
(continued)
2
C (inter-IC bus) serial protocol at a start/stop condition or user defined frame
2 ns (MSO/DSO705xB/710xB scope channels)
2 ns (logic channels on 7000B Series MSO or
MSO-upgraded 7000B Series DSO)
2
C, SPI,