N2735A OPTION 004 AGILENT TECHNOLOGIES, N2735A OPTION 004 Datasheet - Page 21

no-image

N2735A OPTION 004

Manufacturer Part Number
N2735A OPTION 004
Description
Oscilloscope Software
Manufacturer
AGILENT TECHNOLOGIES
Series
DSO7000r
Datasheet

Specifications of N2735A OPTION 004

No. Of Channels
4
Scope Bandwidth
350 MHz
Scope Type
Hand Held
Lead Free Status / RoHS Status
na
Performance characteristics
Trigger system (continued)
Nth edge burst
Autoscale
Scope channel triggering
Range (internal)
Sensitivity*
Coupling
Digital (D15 - D0) channel triggering (7000B Series MSO or MSO-upgraded 7000B Series DSO)
Threshold range (user defined)
Threshold accuracy
Predefined thresholds
External (EXT) triggering
Input impedance
Maximum input
Range
Sensitivity
Coupling
Probe ID
*
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
Trigger on the Nth edge of a burst that occurs after an idle time that you specify. Max edge
count: 65,536.
Finds and displays all active scope and logic (for 7000B Series MSO) channels, sets edge
trigger mode on highest-numbered channel, sets vertical sensitivity on scope channels and
thresholds on logic channels, time base to display ~1.8 periods. Requires minimum voltage
>10 mVpp, 0.5% duty cycle and minimum frequency >50 Hz.
±6 div from center screen
<10 mV/div: greater of 1 div or 5 mV; ≥10 mV/div: 0.6 div
AC, ~10 Hz on MSO/DSO701xB/703xB/705xB/710xB,
DC, noise reject, HF reject and LF reject (~50 kHz)
±8.0 V in 10 mV increments
±(100 mV + 3% of threshold setting)
TTL = 1.4 V, CMOS = 2.5 V, ECL = –1.3 V
MSO/DSO7xx2B (2-/2+16-ch models)
1 MΩ ± 3% || 14 pF or 50 Ω
CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk
With 10073C/D 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50-Ω input
DC coupling: trigger level ±1 V and ±8 V
For ±1 V range setting:
DC to 100 MHz, 100 mV;
For ±8 V range setting:
DC to 100 MHz, 250 mV;
AC (~3.5 Hz), DC, noise reject, HF reject and LF reject (~50 kHz)
Auto probe sense and AutoProbe interface
Agilent- and Tektronix-compatible passive probe sense
>100 MHz to bandwidth of oscilloscope: 500 mV
(continued)
MSO/DSO7xx4B (4-/4+16-ch models)
±15 V
±5 V
DC to 500 MHz: 500 mV
2.14 kΩ ±5%
21