CRT1-TS04T Omron, CRT1-TS04T Datasheet - Page 591

no-image

CRT1-TS04T

Manufacturer Part Number
CRT1-TS04T
Description
CompoNet 4 Pt Thermocouple
Manufacturer
Omron
Datasheet

Specifications of CRT1-TS04T

Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Connecting the Devices
Required for Adjustment
Temperature Input Unit Functions
Note When using an R, S, E, B, or W type thermocouple, a K type can be substi-
Note The personal computer (CX-Integrator) is connected through CompoNet in
■ Reference Voltage/Current Generator and Precision Digital Multimeter
■ Cold Junction Compensator (such as a ZERO-CON 0°C Bath) and
■ CX-Integrator
Compensating Conductors
The following paragraphs explain how to connect the devices that must be
connected to the CRT1-TS04T for user adjustment. Wire the following devices
properly when adjusting the CRT1-TS04T.
Used to make adjustments at the upper limit and lower limit.
Prepare devices that can generate accurate 0 mV, 20 mV, and 50 mV volt-
ages. Use a precision digital multimeter that can measure the output voltage
and indicate when the voltage/current generator is not producing an accurate
voltage output.
Used to adjust the bias compensation value.
The cold junction compensator (the ZERO-CON 0°C bath is used in following
examples) is a device that maintains an accurate 0
thermocouple sensors. Use a cold junction compensator compatible with the
sensor being adjusted.
tuted. Set the ZERO-CON to 0
Actual adjustments are made from the CX-Integrator windows.
Adjustment Device Connection Diagram
Connect the reference voltage/current generator (STV), precision digital multi-
meter (DMM), and ZERO-CON to the input terminals. Connect to the corre-
sponding input terminals when adjusting inputs 0 to 3.
the above diagram. If a CS1W-CRM21 or CJ1W-CRM21 Master Unit is being
used, the Temperature Input Unit can be adjusted by connecting the CX-Inte-
grator through the Master Unit using a peripheral bus connection.
Compensating
conductors
CRT1-TS04T
Reference voltage/current generator (STV)
Remove when performing bias compensation.
Precision digital multimeter (DMM)
Use to confirm that the STV is generating an accurate voltage.
Cold junction compensator (such as a ZERO-CON)
Leave open (disabled) until performing bias compensation.
CompoNet
°
C (32
°
F).
°
C (32
DeviceNet Configurator
CX-Integrator
°
F) temperature for
Section 11-6
567

Related parts for CRT1-TS04T