EM1-H0R0V-1N SENSIRION, EM1-H0R0V-1N Datasheet - Page 12

MASS FLOW METER, FOR GASES

EM1-H0R0V-1N

Manufacturer Part Number
EM1-H0R0V-1N
Description
MASS FLOW METER, FOR GASES
Manufacturer
SENSIRION
Datasheet

Specifications of EM1-H0R0V-1N

Svhc
No SVHC (15-Dec-2010)
Air Flow Range
0 - 0.05 L/min
External Length / Height
104mm
External Width
33.7mm
Height
43.6mm
Length (imperial)
5.35"
Output Configuration
RS232/SPI
Output Type
RS232
Pressure Max
8bar
Supply Voltage Range Dc
7V To 18V
Supply Voltage Dc Max
18VDC
Rohs Compliant
Yes
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
MOSI
SCK
/CS
CMOSens
Figure 10: Timing Diagram of the CMOSens
4.3 SPI Output
The scope diagram further describes the CMOSens
Data on MOSI are valid after falling edge of SCK.
1 = SCK
2 = MOSI
3 = /CS
Figure 11: Scope Diagram. Value: 0000100110010010 bin / 0992 hex / 2450 dec ==> 19.14 ln/min
Note: Free PC Software supporting serial RS232 communication can be downloaded from http://www.sensirion.com
www.sensirion.com
®
EM1 Mass Flow Meter for Gases
Serial clock output
Serial data output
Chip select
tcss
MSB
bit
15
tclk
14...9
bit
®
Data Sheet EM1 - v 2.6 / October 2006
MFM SPI Interface.
®
EM1 device SPI mode data access protocol.
bit
8
tst
twait
bit
7
6...1
bit
LSB
bit
0
12 / 17