5492GPIB B&K Precision, 5492GPIB Datasheet - Page 11

no-image

5492GPIB

Manufacturer Part Number
5492GPIB
Description
DMM BENCH 5 1/2DGT WITH GPIB CRD
Manufacturer
B&K Precision
Type
Digital (DMM)r
Datasheets

Specifications of 5492GPIB

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
120000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Accuracy
+/- 0.012 %
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
!
!
Excessive error may occur when making measurements with 1 to 10 V
resolutions after measuring high voltage up to 1000 volts dc. It requires two
minutes before making low-level measurements.
After making a high current measurement using the 12A input, thermal
voltages are generated that may create errors when making high-resolution
low-level dc measurements of volts, amps, or ohms.
It requires ten minutes to allow the thermals to settle out before making
low-level measurements in order to obtain the best accuracy.
Voltage, Resistance or Frequency Measurements
Press the desired function button and connect the test leads (TL 36 or TL 35
as described in Section 1-3) as shown in Figure 2-5 to measure voltage,
resistance, or frequency. The meter will select the appropriate range in the
auto-range mode, and an annunciator on the display will indicate
measurement units.
Current Measurements
To measure current, connect the test leads to mA input terminal or 12A input
terminal for measured current above 1200mA (5492) / 120mA (5491) as
shown in Figure 2-6.
Be sure to turn off the power in the circuit to be measured before taking
current measurement.
Break the circuit on the groundside to minimize the common mode voltage)
to be measured, and place the meter in series at that point.
Turn on power to the circuit, and then read the display. The meter will select
the appropriate range automatically, and an annunciator on the display will
indicate the units of the measurement value shown.
Turn off power to the circuit and disconnect the meter from the tested circuit.
NOTE
NOTE
11