BK2534 B&K Precision, BK2534 Datasheet - Page 58

OSCOPE 60MHZ DIGITAL STORAGE

BK2534

Manufacturer Part Number
BK2534
Description
OSCOPE 60MHZ DIGITAL STORAGE
Manufacturer
B&K Precision
Type
Bench, Digitalr
Datasheets

Specifications of BK2534

Includes
Probes, Software, USB Cable
Bandwidth
60MHz
Channels
Dual
Display Type
LCD - Color
Features
Software Controllable, USB Port, Waveform Capture
Probe Type
PR-37A x1/x10/Ref Probes (2)
Sampling Rate (per Second)
400MSa/s
Input Impedance
1M - 19pF
Rise Time (typ)
5.83nS
Voltage - Input, Max
400Vpeak
Voltage - Supply
100 ~ 240VAC
Lead Free Status / RoHS Status
Not applicable / Not applicable
Firmware Update
The oscilloscope firmware can be conveniently updated via
the front panel USB host interface.
Press SAVE/LOAD→External Storage to display the
EXTERNAL menu.
Turn the entry knob to select the correct update file. The file
2042A.UPT is selected as shown in the following figure.
Press the Load softkey to start the update operation. A
loading progress bar and then an updating progress bar will
be displayed and indicate the progress of the update
operation.
 
Basic Operation
115
 
116
Basic Operation
Finally, the message “Restart to complete updating” will
be displayed to remind you to restart (power cycle) the
instrument.
If the software update failed, repeat the above procedures to
update again.
Note: The default file extension of the update file is ”.upt”.
Note: Power to the oscilloscope must not be turned off
 
Select the correct update file according to the
model of the oscilloscope. The error message
“Incompatible file ” will be displayed if the file is
not the correct one for your oscilloscope.
during the updating process. If this happens, you
will have to return the instrument to the factory for
service.