CX1594 ETC, CX1594 Datasheet - Page 3

no-image

CX1594

Manufacturer Part Number
CX1594
Description
Deuterium Filled, Hollow Anode, Five-gap Ceramic Thyratron
Manufacturer
ETC
Datasheet
10. In crowbar service, most of the coulombs are often in the
11. Measured with respect to cathode. When grid 1 is pulse
12. DC negative bias voltages must not be applied to grid 1.
13. Typical figures are obtained on test using conditions of
14. The time interval between the instant at which the rising
15. The drift in delay time over a period from 10 seconds to
#E2V Technologies
9. When DC priming is used on grid 1, a negative bias of 100
to 150 V must be applied to grid 2 to ensure anode voltage
hold-off. DC priming is recommended for crowbar service.
power supply follow-on current rather than the storage
capacitor discharge.
driven, the last 0.25 ms of the top of the grid 1 pulse must
overlap the corresponding first 0.25 ms of the top of the
delayed grid 2 pulse. (See schematic diagram, page 5).
When grid 1 is pulse driven, the potential of grid 1 may
vary between 710 and +5 V with respect to cathode
potential during the period between the completion of
recovery and the commencement of the succeeding grid
pulse.
minimum grid 2 drive. Improved performance can be
expected by increasing grid drive.
unloaded grid 2 pulse reaches 25% of its pulse amplitude
and the instant when anode conduction takes place.
10 minutes after reaching full voltage.
HEALTH AND SAFETY HAZARDS
E2V Technologies hydrogen thyratrons are safe to handle and
operate, provided that the relevant precautions stated herein are
observed. E2V Technologies does not accept responsibility for
damage or injury resulting from the use of electronic devices it
produces. Equipment manufacturers and users must ensure that
adequate precautions are taken. Appropriate warning labels and
notices must be provided on equipments incorporating E2V
Technologies devices and in operating manuals.
Equipment must be designed so that personnel cannot come
into contact with high voltage circuits. All high voltage circuits
and terminals must be enclosed and fail-safe interlock switches
must be fitted to disconnect the primary power supply and
discharge all high voltage capacitors and other stored charges
before allowing access. Interlock switches must not be
bypassed to allow operation with access doors open.
All high voltage devices produce X-rays during operation and
may require shielding. The X-ray radiation from hydrogen
thyratrons is usually reduced to a safe level by enclosing the
equipment or shielding the thyratron with at least 1.6 mm
(
Users and equipment manufacturers must check the radiation
level under their maximum operating conditions.
1
/
16
inch) thick steel panels.
High Voltage
X-Ray Radiation
CX1594, page 3

Related parts for CX1594