85RF Fluke Electronics, 85RF Datasheet - Page 18

no-image

85RF

Manufacturer Part Number
85RF
Description
PROBE HIGH FREQ 500MHZ
Manufacturer
Fluke Electronics
Type
High Frequencyr
Datasheets

Specifications of 85RF

Impedance
10M Ohm
Voltage - Max
30VRMS, 200VDC
Lead Free Status / RoHS Status
Lead free by exemption / RoHS compliant by exemption
Ratings
-
Current Rating
-
Color
-
Termination
-
Wire Insulation Material
-
Cable Length
-
Other names
532457
8062A
Instruction Manual
1-6
AC Voltage (True RMS, AC-Coupled)
Ranges .............................................. 200 mV, 2V, 20V, 200V, 750V
Accuracy *.......................................... (% of reading + no. of digits). See table
199.99 mV
* Not specified for input < 10% of range.
1.9999V
19.999V
199.99V
Input Impedance ......................... 10 M shunted by <100 pF
Common Mode Noise
Rejection ..................................... >60 dB at 50 Hz and 60 Hz (1 k
Crest Factor Range..................... 1:1 to 3:1
Response Time........................... Five seconds maximum to rated
Overload Protection .................... 750V rms or 1000V peak continuous
Voltage
.2000 -
2.000 -
20.00 -
499.9V
500.0 -
750.0V
20.0 -
75.0 -
Input
Resolution
100 mV
.01 mV
10 mV
.1 mV
1 mV
Table 1-2. 8062A Specifications (cont)
200 mV
Range
200V
750V
20V
2V
Specified
1% +10
20 Hz -
45 Hz
Not
imbalance)
accuracy within selected range, 12
seconds to rated accuracy from an
overload.
except 20 seconds maximum on the
200 mV range above 300V rms or
300V dc. Input not to exceed a volt-
hertz product of 1x10
200V at 50 kHz).
below:
0.5% +12
45 Hz -
500 Hz
+10
+10
1%
2%
1 kHz
500 Hz -
5% +20
10 kHz
0.5%
+ 20
Not Specified
7
(for example,
10 kHz -
5% +40
30 kHz
+ 40
1%