Since average-responding meters have been in use for so long, you may have
accumulated test or reference data based on them. The conversion factors in
Figure 2-8 should help you convert between the two measurement methods.
2-15. High Impedance DC Voltage
Occasionally you may want to make dc voltage measurements in high
impedance circuitry where even the 10 M input impedance for the normal
dc voltage function could load the circuit and cause significant errors. For
example, a 10 M input impedance causes a .1% error when measuring the
voltage across the 10 k leg of a 90 k over 10 k voltage divider. The
8062A offers a >1000 M (typically >10,000 M ) input impedance dc
voltage function which greatly reduces this error.
Figure 2-9 describes how to select the high input impedance dc voltage
function (the ac voltage function does not operate in this mode). Notice that
all of the function switches must be out to select this function. Either the 2V
or the 200 mV ranges may be selected.
2-14
High Impedance
DC Voltage (V)
1. Select the 2V or
2. Ensure all function
3. Connect the test leads as shown.
4. Heed the input overload limits
5. Read the measured value on the display.
the 200 mV range.
switches are out.
(Table 2-2) and connect the leads to
the circuit being measured.
Figure 2-9. High Impedance DC Voltage
Low (-)
High (+)
dy10f.eps