4-20. Troubleshooting
If necessary, refer to Chapter 2 for operating instructions or Chapter 3 for the
theory of operation. The troubleshooting information is supported by the
schematics and tables in Chapter 7.
4-21. Self-Tests
The 8062A offers three self-tests: power-on self-test, ratio self-test, and
switch decoding self-test. The power-on self-test is automatically performed
whenever the instrument is turned on. It is described in Chapters 2 and 3.
The other two tests function as follows:
4-22. Ratio Self-Test
The ratio self-test is an operating mode of the 8062A in which the reference
voltage for the a/d converter is applied to the a/d converter during both the
integrate and the read periods. If the instrument is functioning properly, the
display should read 10000
on the range, and does not affect the number of counts).
To select the ratio self-test, select a voltage or current function. Hold down
the
has been completed (the display is .8.8.8.8) release the
instrument should now be in the ratio self-test mode. To cancel the ratio self-
test, press the
If the count is within tolerance, it gives a strong indication that the a/d
converter is working properly. If the count deviates more than 5 counts from
10000, the probable causes are as follows ( in order of probability): a/d
converter in U3, leakage around or failure of C16, C18, Z3, R8, or the power
supply.
Static discharge can damage MOS components U3, U4,
and U5. Follow the handling precautions for static
sensitive devices previously described in this chapter.
Never remove, install or otherwise connect or disconnect
components without first setting the instrument power
switch off and disconnecting any inputs to the
instrument.
button while you turn on the instrument. After the power-self-test
button or turn off the instrument.
10 counts (the decimal point location depends
Caution
Troubleshooting
Maintenance
button. The