4-6.
Use the following procedure to gain access to the calibration adjustments or
the backup fuse (F2):
1.
2.
3.
4.
4-6
It is not necessary to remove the main pcb from the bottom case in
order to disassemble or reassemble the LCD. However, because the
LCD and the microcomputer require similar special handling, the
disassembly and reassembly procedures are described together.
To avoid contaminating the pcbs with oil from the
fingers, handle the pcbs by the edges or wear gloves.
Disconnect the test leads and battery eliminator, if attached. Turn the
power switch off.
Remove the three phillips screws from the bottom of the case.
Turn the instrument face-up and grasp the top cover at both sides of the
input connectors. Then pull the top cover from the unit. The backup fuse
and the calibration adjustments are now accessible (Figure 4-1).
The function buttons below the display are part of a
single elastomeric strip (Figure 4-1) that is held in place
by the top cover. When the top cover is removed, the
elastomeric strip will be loose and may be removed. Do
not touch or contaminate the carbon-impregnated switch
contacts on the bottom of the strip or the switch
contacts on the microcomputer pcb. If the contacts do
become contaminated, clean them with isopropyl
alcohol.
To reassemble, position the elastomeric strip on the microcomputer pcb
so that the small rubber posts on the bottom of the strip are properly
seated. Install the top cover and fasten the three screws on the bottom
case.