MPXC2011DT1 Freescale Semiconductor, MPXC2011DT1 Datasheet - Page 3

SENSOR PRESSURE MEDICAL CHIP PAK

MPXC2011DT1

Manufacturer Part Number
MPXC2011DT1
Description
SENSOR PRESSURE MEDICAL CHIP PAK
Manufacturer
Freescale Semiconductor
Series
MPXC2011r
Datasheet

Specifications of MPXC2011DT1

Pressure Type
Differential
Operating Pressure
1.45 PSI, 10 kPa
Port Size
Female, 0.128" (3.2512mm)
Output
24 ~ 26mV
Voltage - Supply
10V
Termination Style
PCB
Operating Temperature
15°C ~ 40°C
Package / Case
4-ChipPak
Mounting Style
Through Hole
Maximum Operating Temperature
+ 40 C
Minimum Operating Temperature
+ 15 C
Operating Supply Voltage
3 V
Lead Free Status / RoHS Status
Lead free / RoHS Compliant

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
MPXC2011DT1
Manufacturer:
FREESCALE
Quantity:
20 000
Sensors
Freescale Semiconductor
MAXIMUM RATINGS
OPERATING CHARACTERISTICS
Table 2. Operating Characteristics (V
Table 1. Maximum Ratings
Pressure Range
Supply Voltage
Supply Current
Full Scale Span
Offset
Sensitivity
Linearity
Pressure Hysteresis
Temperature Hysteresis
Temperature Effect on Full Scale Span
Temperature Effect on Offset
Input Impedance
Output Impedance
Response Time
Warm-Up
Offset Stability
1. 1.0 kPa (kiloPascal) equals 0.145 psi.
2. Device is ratiometric within this specified excitation range. Operating the device above the specified excitation range may induce additional
3. Full Scale Span (V
4. Offset (V
5. Accuracy (error budget) consists of the following:
6. Response Time is defined as the time for the incremental change in the output to go from 10% to 90% of its final value when subjected to a
7. Offset stability is the product's output deviation when subjected to 1000 hours of Pulsed Pressure, Temperature Cycling with Bias Test.
Maximum Pressure (Backside)
Storage Temperature
Operating Temperature
1. Exposure beyond the specified limits may cause permanent damage or degradation to the device.
error due to device self-heating.
minimum rated pressure.
Linearity:
Temperature Hysteresis:
Pressure Hysteresis:
TcSpan:
TcOffset:
specified step change in pressure.
(4)
(5)
off
) is defined as the output voltage at the minimum rated pressure.
(7)
(2)
(3)
(6)
(1)
(10% to 90%)
(5)
FSS
(0 to 10 kPa)
(5)
) is defined as the algebraic difference between the output voltage at full rated pressure and the output voltage at the
(+15°C to +40°C)
Characteristic
(5)
Rating
Output deviation from a straight line relationship with pressure, using end point method, over the specified
pressure range.
Output deviation at any temperature within the operating temperature range, after the temperature is cycled to
and from the minimum or maximum operating temperature points, with zero differential pressure applied.
minimum or maximum rated pressure, at 25°C.
Output deviation at full rated pressure over the temperature range of 0 to 85°C, relative to 25°C.
Output deviation with minimum rated pressure applied, over the temperature range of 0 to 85°C, relative to 25°C.
Output deviation at any pressure within the specified range, when this pressure is cycled to and from the
(1)
(5)
S
= 10 Vdc, T
A
= 25°C unless otherwise noted, P1 > P2)
Symbol
Symbol
TCV
TCV
ΔV/ΔP
P
V
T
P
Z
T
V
max
V
Z
FSS
t
stg
I
OP
out
A
R
o
off
in
S
FSS
off
1300
1400
Min
-1.0
-1.0
-1.0
-1.0
24
0
+15 to +40
-25 to +85
Value
75
±0.1
±0.1
±0.5
Typ
6.0
2.5
1.0
25
20
3
2550
3000
Max
1.0
1.0
1.0
1.0
10
10
26
v
Pressure
MPXC2011DT1
Unit
kPa
°C
°C
mV/kPa
%V
%V
%V
%V
%V
mAdc
Unit
kPa
Vdc
mV
mV
mV
ms
ms
Ω
Ω
FSS
FSS
FSS
FSS
FSS
3

Related parts for MPXC2011DT1